实时三维超声心动图估测左心室容量的实验研究
引用本文:陈国珍,孙锟,陈树宝,张玉奇,洪雯静,薛海红. 实时三维超声心动图估测左心室容量的实验研究[J]. 中国医学影像技术, 2004, 20(6): 844-846
作者姓名:陈国珍  孙锟  陈树宝  张玉奇  洪雯静  薛海红
作者单位:上海第二医科大学附属新华医院上海儿童医学中心心内科,上海,200127
摘    要:目的初探实时三维超声心动图(3DE)估测心室容量的可行性和准确性.方法应用Philips公司实时3DE系统采集19只离体猪心金字塔形数据库,结合相应测量软件用心尖长轴系列平面法分别测量左室容量,并与二维双平面Simpson法和猪心左室排水法实测值比较.结果从方差分析、SNK检验及直线相关分析看:实时3DE 16平面法(23.31±14.29)ml、8平面法(23.42±14.56)ml估测左室容量值与猪心左室排水法实测值(23.79±14.78)ml差异无统计学意义(P>0.05),实时3DE 2平面法(17.94±10.46)ml和二维超声双平面Simpson法(19.67±13.46)m1测值与排水法实测值差异则有统计学意义(P<0.05).而且实时3DE 16平面和8平面法测值与排水法实测值高度相关(r=0.98,P<0.0001),二维超声双平面Simpson法则较逊(r=0.89).结论实时3DE是心室容量准确估测的又一可靠手段.

关 键 词:超声心动图描计术,三维  心室容量,左  动物,实验
文章编号:1003-3289(2004)06-0844-03
修稿时间:2003-12-16

Initial study of real-time three-dimensional echocardiography for quantifying left ventricular volume in vitro
CHEN Guo-zhen,SUN Kun,CHEN Shu-bao,ZHANG Yu-qi,HONG Wen-jing,and XUE Hai-hong. Initial study of real-time three-dimensional echocardiography for quantifying left ventricular volume in vitro[J]. Chinese Journal of Medical Imaging Technology, 2004, 20(6): 844-846
Authors:CHEN Guo-zhen  SUN Kun  CHEN Shu-bao  ZHANG Yu-qi  HONG Wen-jing  and XUE Hai-hong
Affiliation:Department of Pediatric Cardiology, Shanghai Children's Medical Center, Xinhua Hospital, Shanghai Second Medical University, Shanghai 200127, China;Department of Pediatric Cardiology, Shanghai Children's Medical Center, Xinhua Hospital, Shanghai Second Medical University, Shanghai 200127, China;Department of Pediatric Cardiology, Shanghai Children's Medical Center, Xinhua Hospital, Shanghai Second Medical University, Shanghai 200127, China;Department of Pediatric Cardiology, Shanghai Children's Medical Center, Xinhua Hospital, Shanghai Second Medical University, Shanghai 200127, China;Department of Pediatric Cardiology, Shanghai Children's Medical Center, Xinhua Hospital, Shanghai Second Medical University, Shanghai 200127, China;Department of Pediatric Cardiology, Shanghai Children's Medical Center, Xinhua Hospital, Shanghai Second Medical University, Shanghai 200127, China
Abstract:Objective To evaluate initially the feasibility and accuracy of real-time three-dimensional echocardiography (3DE) for quantifying left ventricular volume in an in vitro experimental study. Methods Nineteen excised porcine hearts were examined by real-time 3DE and conventional two-dimensional (2D) echocardiography with Philips SONOS 7500 echocardiographic system. The measurement of their left ventricular volume examined by real-time 3DE was outlined and calculated by 2-plane, 4-plane, 8-plane and 16-plane method with the 4D Cardio-View RT of TOMTEC image system, and their results were compared with those of 2D biplane Simpson-method. The true volume of excised porcine left ventricle measured by water displacement was served as a reference standard. Results "Full volume" imaging in all excised porcine hearts was well displayed by real-time 3DE. There was no significant difference (P>0.05) between the measurements of real-time 3DE 16-plane (23.31±14.29) ml, 8-plane (23.42±14.56) ml and those of true volume of excised porcine left ventricle (23.79±14.78) ml, but there was significant difference (P<0.05) between the measurements of real-time 3DE 2-plane (17.94±10.46) ml, 2D biplane Simpson-method (19.67±13.46) ml and those of true left ventricular volume. Meanwhile,the measurements of RT-3DE 16-plane and 8-plane were high closer to those of true left ventricular volume. The correlation of real-time 3DE and 8-plane (r=0.98, P<0.0001) with true left ventricular volume was more than that of 2DE biplane Simpson-method (r=0.89). Conclusion real-time 3DE will be a valuable and accurate technique to quantify ventricular volume.
Keywords:Echocardiography   three-dimensional  Ventricular volume   left  Animals   laboratory