Development of an Autoflow Cruise Control System for a Centrifugal Pump
Authors:Hiroshi Nishida  Toshiyuki Beppu  Masato Nakajima  Tomohiro Nishinaka  Hayao Nakatani  Kenji Ihashi  Takahiro Katsumata  Masaya Kitamura  Shigeyuki Aomi  Masahiro Endo  Hitoshi Koyanagi  Hiroaki Oshiyama  Atsuhiko Nogawa  Nobuhiko Amano  Chisato Nojiri  Tetsuzo Akutsu
Affiliation:Department of Cardiovascular Surgery, The Heart Institute of Japan, Tokyo Women's Medical College, Tokyo, Japan;Terumo Corporation, Tokyo, Japan
Abstract:Abstract: To improve the ease of driving a centrifugal pump that is afterload dependent, we have developed an automatic flow control system for the Terumo Capiox centrifugal pump system. This system consists of an autoflow cruise control system with a safety cutoff. The Capiox Pump Console 3000 was controlled by a personal computer through a serial communication line. In the usual manual mode, the motor speed knob works as a pump speed control, and in the autoflow mode, the same knob works as a blood flow rate control. After selecting and obtaining the desired flow rate, the mode was changed from manual to autoflow mode. In the autoflow mode, the computer compares the desired flow rate with the actual flow measured by an ultrasonic Doppler flow-meter and adjusts the motor rotational speed accordingly. During both in vivo and in vitro testing, this autoflow mode was able to return the changed flow that was disrupted by either clamping and declamping of the tubing or by the bolus injection of a vasomotor drug to the selected flow rate within 10 s without any significant fluctuation. In conclusion, the newly developed computer controlled autoflow system was able to produce a reliable and effective flow regulation for a centrifugal pump.
Keywords:Centrifugal pump    Autoflow control    Ultrasonic Doppler flowmeter
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号