Dosimetric Characteristics of the Siemens IGRT Carbon Fiber Tabletop
Authors:Emiliano Spezi PhD  Andrea Ferri  
Institution:aServizio di Fisica Sanitaria, Policlinico S. Orsola-Malpighi, Bologna, Italy
Abstract:In this work, the dosimetric characteristics of a new commercial carbon fiber treatment table are investigated. The photon beam attenuation properties of the Siemens image-guided radiation therapy (IGRT) tabletop were studied in detail. Two sets of dosimetric measurements were performed. In the first experiment a polystyrene slab phantom was used: the central axis attenuation and the skin-sparing detriment were investigated. In the second experiment, the off-axis treatment table transmission was investigated using a polystyrene cylindrical phantom. Measurements were taken at the isocenter for a 360° rotation of the radiation beam. Our results show that the photon beam attenuation of the Siemens IGRT carbon fiber tabletop varies from a minimum of 2.1% (central axis) to a maximum of 4.6% (120° and 240° beam incidence). The beam entrance dose increases from 82% to 97% of the dose at the depth of maximum for a clinical 6-MV radiation field. The depth of maximum also decreases by 0.4 cm. Despite the wedge cross section of the table the beam attenuation properties of the IGRT tabletop remain constant along the longitudinal direction. American Association of Medical Dosimetrists.
Keywords:Radiotherapy Treatment table  IMRT  IGRT  PACS