Comparison of Dioptron Nova refractive data with conventional refractive data
Authors:D M Perrigin  T Grosvenor  A Reis  J Perrigin
Abstract:Refractive data obtained by means of the Dioptron Nova Diagnostic Eye Computer were compared to data obtained by conventional clinical refraction on a group of 236 clinic patients. Spherical equivalent power data were found to average -0.32 D more for Dioptron Nova data than for clinical data, whereas spherical power data averaged -0.25 D more for Dioptron Nova data than for clinical data, and cylinder power averaged -0.12 D more for Dioptron Nova data. Dioptron and clinical data were found to be within +/- 0.50 D of each other for 74% of eyes for spherical equivalent power, for 83% of eyes for spherical power, and for 91% of eyes for cylinder power. Cylinder axis data were found to be within +/- 10 degrees for the two methods for 78% of eyes.