Initial Experience with a New Balloon-Guided Single Lead Catheter for Internal Cardioversion of Atrial Fibrillation and Dual Chamber Pacing
Authors:A. PLEWAN  C. VALINA  R. HERRMANN  E. ALT
Affiliation:I Medizinische Klinik, Klinikum rechts der Isar, München, Germany
Abstract:Background: Based on the observation that internal cardioversion (IntCV) of atrial fibrillation is effective with electrodes in the right atrium and pulmonary artery, a new balloon-guided catheter and external defibrillation device with optional dual chamber pacing was evaluated. Methods: IntCV was attempted in 27 patients (age: 57 ± 10 years, duration: 14 ± 18 months, left atrial diameter 56 ± 8 mm) using a new defibrillation device (Alert, EP MedSystems, Inc., NJ, USA) that allows the delivery of biphasic shocks (0.5–15 J, variable tilt), atrial and ventricular pacing, and online signal recording. Pacing and defibrillation shocks were applied via a 7.5 Fr balloon-guided catheter (EP MedSystems, Inc.). Pacing, sensing, and triggering were established through the proximal atrial array and an electrode ring between both defibrillation arrays and a single ventricular electrode ring. Catheters were inserted from the antecubital vein. Results: In 25 of 27 patients sinus rhythm was restored with a mean energy of 6.7 ± 4.5 J. In five patients, atrial postshock pacing was required for bradycardia and atrial premature beats. The mean fluoroscopy time was 2.0 ± 1.3 minutes. Conclusion: The high success rate, ease of application, and backup dual chamber pacing suggest that this system is an alternative to established methods of cardioversion. In certain indications, such as failure of prior external cardioversion and situations in which a standard pulmonary balloon catheter is needed, this system would be advantageous.
Keywords:atrial fibrillation    internal cardioversion    defibrillation system    dual chamber pacing
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号