首页 | 本学科首页   官方微博 | 高级检索  
检索        


Self-expanding stents in transjugular intrahepatic portosystemic shunt: experience with nitinol Strecker stents
Authors:P Rossi  M Bezzi  F M Salvatori  L Broglia  F Maccioni  G Pizzi  S Abbondanza  G Bonomo
Institution:(1) Department of Radiology, University of Rome ldquoLa Sapienzardquo Policlinico Umberto I, Viale Regina Elena 324, I-00161 Rome, Italy
Abstract:The aim of our study was to evaluate the performance and efficacy of a new self-expanding stent (nitinol Strecker stent) in the transjugular intrahepatic portosystemic shunt (TIPS) procedure. We have successfully placed 64 nitinol Strecker stents in 48 patients. The average portosystemic gradient decreased from 22 to 11 mm Hg. Balloon dilatation was necessary in 12 of 35 angiographically controlled cases at 5 days (34%), because of incomplete stent expansion, small thrombi within the stent or obstruction. At 1–6 months stent malfunctions occurred in 8 of 23 patients who underwent control angiography (34%) and at 6–24 months in 6 of 7 patients (85%). Rebleeding occurred in 2 of 39 patients (follow-up > 1 month) (5%) and temporary crises of de novo encephalopathy were observed in 11 of 48 patients (23%). Refractory ascites completely resolved in 4 of 6 patients (66%) and improved in the remaining 2 cases. Compared with other self-expanding stents, nitinol Strecker stents seem to be equally effective in TIPS; no increase in complication rate was observed, either clinical or stent-related. Correspondence to: P. Rossi
Keywords:Transjugular intrahepatic portosystemic shunt (TIPS)  Nitinol Strecker stents  Metallic stent  Interventional radiology
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏