Effects of radiofrequency pulses delivered in the vicinity of the coronary arteries: implications for nonsurgical transthoracic epicardial catheter ablation to treat ventricular tachycardia
Authors:D'Avila André  Gutierrez Paulo  Scanavacca Mauricio  Reddy Vivek  Lustgarten Daniel L  Sosa Eduardo  Ramires José Antonio F
Affiliation:Heart Institute (InCor), University of S?o Paulo Medical School, S?o Paulo, Brazil. andre.davila@incor.usp.br
Abstract:This study evaluates the effects of epicardial RF ablation on the coronary vessels in a canine model. Nonsurgical epicardial catheter ablation is a minimally invasive procedure that has proven to be efficacious for the treatment of VT. This approach is limited by concern regarding the potential adverse effects of RF ablation on the epicardial coronary arteries. After lateral thoracotomy, a multipolar linear ablation catheter was sewn adjacent to or crossing the LAD in nine mongrel dogs. Five of these dogs also had a standard 4-mm tip ablation catheter sewn adjacent to the LAD. RF pulses were delivered using temperature control. Animals were sacrificed 14 days after ablation for histological analyses. Using the 4-mm tip catheter, 22 consecutive single lesions of 2.67 +/- 0.35 mm depth were studied. The only coronary arterial change seen was replacement of the media with extracellular matrix. Using the linear ablation catheter, 117 pulses were applied to generate 24 linear lesions (3.8 +/- 1 mm depth). Replacement of the coronary arterial media with extracellular matrix proliferation was commonly seen, and severe hyperplasia occurred in one artery. Intravascular thrombosis occurred in six arteries. The internal perimeter of the vessel (0.78 +/- 0.49 mm vs 1.79 +/- 0.83 mm) was the only variable associated with severe arterial damage. The effects of RF ablation delivered adjacent to the LAD were limited to the media but when delivered above the artery, severe intimal hyperplasia and intravascular thrombosis may occur. Susceptibility to damage is inversely proportional to the vessel size.
Keywords:catheter ablation    coronary artery    arrhythmia    epicardium    ventricular tachycardia
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号