首页 | 本学科首页   官方微博 | 高级检索  
     


Monodisperse Versus Polydisperse Ultrasound Contrast Agents: Non-Linear Response,Sensitivity, and Deep Tissue Imaging Potential
Authors:Tim Segers  Pieter Kruizinga  Maarten P. Kok  Guillaume Lajoinie  Nico de Jong  Michel Versluis
Abstract:It has been proposed that monodisperse microbubble ultrasound contrast agents further increase the signal-to-noise ratio of contrast-enhanced ultrasound imaging. Here, the sensitivity of a polydisperse pre-clinical agent was compared experimentally with that of its size- and acoustically sorted derivatives by using narrowband pressure- and frequency-dependent scattering and attenuation measurements. The sorted monodisperse agents had up to a two-orders-of-magnitude increase in sensitivity, that is, in the average scattering cross section per bubble. Moreover, we found, for the first time, that the highly non-linear response of acoustically sorted microbubbles can be exploited to confine scattering and attenuation to the focal region of ultrasound fields used in clinical imaging. This property is a result of minimal pre-focal scattering and attenuation and can be used to minimize shadowing effects in deep tissue imaging. Moreover, it potentially allows for more localized therapy using microbubbles through the spatial control of resonant microbubble oscillations.
Keywords:Monodisperse bubbles  Ultrasound contrast agents  Non-linear echo
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号