24小时食管pH监测电极的三种定位方法比较及其影响因素
引用本文:夏志伟,邸岩,葛颖,王爱英,刘欣,张耀鹏,王琨,段丽萍. 24小时食管pH监测电极的三种定位方法比较及其影响因素[J]. 中国医师进修杂志, 2010, 33(34). DOI: 10.3760/cma.j.issn.1673-4904.2010.34.006
作者姓名:夏志伟  邸岩  葛颖  王爱英  刘欣  张耀鹏  王琨  段丽萍
作者单位:北京大学第三医院消化科,100191
摘    要:目的 探讨食管测压法(M)、pH梯度定位法(P)及X线透视定位法(X)定位24 h食管pH监测电极的差别及其影响因素.方法 50例患者经M测定下食管括约肌距鼻外缘的长度,记录P定位的pH电极位置距鼻外缘的长度,在X线透视下观察电极位置与胃底贲门交界的距离.以差值绝对值≤2 cm为符合标准,分析三种定位方法的差异,并分析影响差别的相关因素.结果 经M、P、X定位的电极定位点距鼻外缘的长度分别为(37.9±2.9)、(40.8±4.4)和(40.7±5.3)cm,P与M以及X与M比较差异均有统计学意义(P<0.01).P与M符合率62.0%(31/50);P与X符合率84.0%(42/50);X与M符合率58.0%(29/50).P较M偏向胃侧(2.9±3.9)cm.影响P与M差别的主要因素为年龄、身高和下食管括约肌的长度(P<0.01或<0.05).影响X和M的因素为身高和下食管括约肌长度(P<0.05或<0.01).结论 与M比较,P确定的位置倾向于偏向胃侧,在身材较高、下食管括约肌长度较长的受试者中易出现差异.X的变异不易确定,亦受身高的影响.

关 键 词:食管  氢离子浓度  电极

Differences of three positioning methods of the 24-hour esophageal pH catheter and relative factors
XIA Zhi-wei,DI Yan,GE Ying,WANG Ai-ying,LIU Xin,ZHANG Yao-peng,WANG Kun,DUAN Li-ping. Differences of three positioning methods of the 24-hour esophageal pH catheter and relative factors[J]. Chinese Journal of Postgraduates of Medicine, 2010, 33(34). DOI: 10.3760/cma.j.issn.1673-4904.2010.34.006
Authors:XIA Zhi-wei  DI Yan  GE Ying  WANG Ai-ying  LIU Xin  ZHANG Yao-peng  WANG Kun  DUAN Li-ping
Abstract:Objective To study the differences of esophageal manometry (M),pH set-up method (P) and X-ray (X) on positioning of the 24-hour esophageal pH catheter and relative factors. Methods Fifty subjects underwent M, then pH catheter was located by P and X. The difference between methods and the relative factors such as body height,lower esophageal sphincter (LES) length etc were analyzed. The difference of less than 2 cm between two methods was accepted. Results The length between the location determined by M,Pand X and nose were (37.9 ±2.9),(40.8 ±4.4) and (40.7 ±5.3) cm, respectively.There was significant difference between P and M as well as between X and M (P< 0.01 ). The coincidence rate was 62.0%(31/50) between P and M;84.0%(42/50) between P and X;and 58.0%(29/50) between X and M. Compared with P,M was (2.9 ±3.9) cm nearer to the stomach. Age, body height and LES length were main factors which affect the difference between P and M (P< 0.01 or < 0.05),body height and LES length were main factors which affect the difference between X and M (P<0.05 or <0.01). Conclusions Compared with M, the location determined by P is nearer to stomach. The location of X is varied. Body height and LES length are main relative factors.
Keywords:Esophagus  Hydrogen-ion concentration  Electrodes