首页 | 本学科首页   官方微博 | 高级检索  
检索        


New MR imaging methods for metallic implants in the knee: artifact correction and clinical impact
Authors:Chen Christina A  Chen Weitian  Goodman Stuart B  Hargreaves Brian A  Koch Kevin M  Lu Wenmaio  Brau Anja C  Draper Christine E  Delp Scott L  Gold Garry E
Institution:Department of Radiology, Stanford University, Stanford, California, USA. tinac@stanford.edu
Abstract:

Purpose:

To evaluate two magnetic resonance imaging (MRI) techniques, slice encoding for metal artifact correction (SEMAC) and multiacquisition variable‐resonance image combination (MAVRIC), for their ability to correct for artifacts in postoperative knees with metal.

Materials and Methods:

A total of 25 knees were imaged in this study. Fourteen total knee replacements (TKRs) in volunteers were scanned with SEMAC, MAVRIC, and 2D fast spin‐echo (FSE) to measure artifact extent and implant rotation. The ability of the sequences to measure implant rotation and dimensions was compared in a TKR knee model. Eleven patients with a variety of metallic hardware were imaged with SEMAC and FSE to compare artifact extent and subsequent patient management was recorded.

Results:

SEMAC and MAVRIC significantly reduced artifact extent compared to FSE (P < 0.0001) and were similar to each other (P = 0.58), allowing accurate measurement of implant dimensions and rotation. The TKRs were properly aligned in the volunteers. Clinical imaging with SEMAC in symptomatic knees significantly reduced artifact (P < 0.05) and showed findings that were on the majority confirmed by subsequent noninvasive or invasive patient studies.

Conclusion:

SEMAC and MAVRIC correct for metal artifact, noninvasively providing high‐resolution images with superb bone and soft tissue contrast. J. Magn. Reson. Imaging 2011;33:1121–1127. © 2011 Wiley‐Liss, Inc.
Keywords:magnetic resonance imaging  metal artifact  SEMAC  MAVRIC  total knee replacement