Measurement of B-type natriuretic peptide by two assays utilizing antibodies with different epitope specificity
Authors:Tamm Natalia N  Semenov Alexander G  Seferian Karina R  Bereznikova Anastasia V  Murakami Maryann M  Apple Fred S  Koshkina Ekaterina V  Krasnoselsky Mihail I  Katrukha Alexey G
Institution:HyTest LTD, Intelligate, 6th floor, Joukahaisenkatu 6, 20520 Turku, Finland.
Abstract:ObjectivesTo compare plasma BNP values determined by a conventional-type and a novel-type of BNP assays.Design and MethodsPlasma samples (n = 94) from HF patients were analyzed by the novel-type “Single Epitope Sandwich” (SES assay prototype) and the conventional-type Siemens ADVIA Centaur BNP assays. Both assays were calibrated using recombinant proBNP (expressed in E. coli).ResultsThe SES assay measured 1.2- to 7.2-fold (2.1 ± 0.9; mean, SD) greater BNP concentrations compared to the Siemens assay. A subset of six samples (6.4%) demonstrated the largest (3- to 7.2-fold higher) difference.ConclusionsThe SES prototype assay appears to more accurately measure the absolute concentrations of BNP immunoreactive forms.