首页 | 本学科首页   官方微博 | 高级检索  
检索        


Catheter closure of accessory pulmonary blood flow after bidirectional Glenn anastomosis using Amplatzer duct occluder.
Authors:Makram R Ebeid  Charles H Gaymes  James A Joransen
Institution:Division of Cardiology, Department of Pediatrics, University Of Mississippi Medical Center, Jackson, Mississippi, 39216, USA. mebeid@ped.umsmed.edu
Abstract:Additional source of pulmonary blood flow in patients with bidirectional cavopulmonary anastomosis (Glenn shunt) may cause elevation of the pulmonary artery pressure precluding safe completion of the Fontan operation. A case is presented with single-ventricle bidirectional cavopulmonary anastomosis and additional flow from the ventricle to the pulmonary artery resulting in elevated Glenn pressure. The communication was successfully occluded using Amplatzer duct occluder with satisfactory reduction in the Glenn pressure.
Keywords:catheter intervention  single ventricle  Glenn  accessory pulmonary flow
设为首页 | 免责声明 | 关于勤云 | 加入收藏