几种血细胞分析仪结果的比对和质控
引用本文:彭黎明,李丽娟,彭志勇.几种血细胞分析仪结果的比对和质控[J].中华检验医学杂志,2000,23(2):94-97.
作者姓名:彭黎明  李丽娟  彭志勇
作者单位:[1]华西医科大学附属第一医院检验科 [2]山东省东营市胜利油田胜利医院检验科
摘    要:目的 为了对不同血细胞分析仪结果进行比对和质量控制。方法 每天随机选取高中、低值病人8例,共分析6d。先将NE-1500与手工法计数比对,再以NE-1500评价本室其他5台血细胞分析仪结果「WBC、RBC、血红蛋白(HGB)、红细胞比积(HCT)」,同时还用新鲜全血对各台血细胞分析仪进行质控「WBC、RBC、HGB、HCT、平均红细胞容积(MCV)、血小板(PLT)」。结果 新鲜全血的质控结果显示

关 键 词:血细胞计数  质量控制  血细胞分析仪  HGB  HCT
修稿时间:1999-08-20

Comparison of the results produced by different hemotology analyzers
PENG Liming,LI Lijuan,PENG Zhiyong,et al..Comparison of the results produced by different hemotology analyzers[J].Chinese Journal of Laboratory Medicine,2000,23(2):94-97.
Authors:PENG Liming  LI Lijuan  PENG Zhiyong  
Institution:PENG Liming*,LI Lijuan,PENG Zhiyong,et al. *Department of Laboratory Medicine,First University Hospital,West China University of Medical Sciences,Chengdu 610041,China
Abstract:Objective To compare the results obtained from different hematology analyzers and set up quality control program for complete blood analysis. Method Eight samples of daily whole blood of high, normal and low level values in patients were collected and analysed for six days. NE 1500 was compared with manual analysis of WBC, RBC, HGB and HCT. NE 1500 was used as a reference analyzer to evaluate the other five hematology analyzers with the same four indicators. Fresh blood was used to calibrate all the six analyzers with WBC, RBC, HGB, HCT, MCV and PLT. Results The coefficient of variance ( CV ) of all the indicators but PLT was smaller than 5%. The CV of PLT was about 10%. The relationship between the results of NE 1500 and manual count was close ( r >0 975). COULTER ONYX, CD 1600(1) and NE 1500 demonstrated a good correlation ( r >0 975); CD 1600(2) and NE 1500 had some discrepancies in results. But HGB and HCT were good ( r >0 975), and the other two, WBC and RBC, had significant discrepancies ( r <0 975). F 820 and NE 1500 showed poor agreement in their results, but one indicator showed a good correlation ( r >0 975), the rest were poor ( r <0 975). Conclusion It is important for laboratories to compare the results of all the different hematology analyzers and to carry out quality control for results of hematology analyzers with fresh blood every day, which is also necessary to ensure the accuracy and precision of the results.
Keywords:Blood  Blood cell count  Quality control