首页 | 本学科首页   官方微博 | 高级检索  
检索        

日立7170和贝克曼DXC800全自动生化分析仪天门冬氨酸氨基转移酶检测结果比对评价
引用本文:许坚锋,钟志娟,郭国威,戴小波,唐文志.日立7170和贝克曼DXC800全自动生化分析仪天门冬氨酸氨基转移酶检测结果比对评价[J].检验医学与临床,2014,0(4):455-456.
作者姓名:许坚锋  钟志娟  郭国威  戴小波  唐文志
作者单位:许坚锋 (广东省中医院珠海医院,广东珠海,519015); 钟志娟 (中山大学附属第五医院,广东珠海,519000); 郭国威 (广东省中医院珠海医院,广东珠海,519015); 戴小波 (广东省中医院珠海医院,广东珠海,519015); 唐文志 (广东省中医院珠海医院,广东珠海,519015);
摘    要:目的:探讨日立7170全自动生化分析仪(简称7170分析仪)和贝克曼 DXC800全自动生化分析仪(检测DXC800分析仪)检测天门冬氨酸氨基转移酶(AST)结果的可比性。方法以7170分析仪作为参考仪器, DXC800分析仪作为实验仪器,按照美国临床和实验室标准化协会(CLSI)EP9-A2文件的要求设计比对方案,以采集自临床患者的空腹新鲜血清标本为待测标本,以美国临床实验室修正法规1988(CLIA′88)允许总误差的1/2作为标准,对两台仪器检测结果间偏差进行评估。结果两台仪器 AST检测结果有较好的相关性,结果间的偏差在临床可接受范围内。结论7170分析仪和DXC800分析仪检测 AST的结果具有可比性,且一致性较好。

关 键 词:仪器比对  天门冬氨酸氨基转移酶  日立7170全自动生化分析仪  贝克曼DXC800全自动生化分析仪

Alignment evaluation of aspartate aminotransferase detected results between two full automatic biochemical analyzers
XUJian-feng,ZHONG Zhi-juan,GUO Guo-wei,DAI Xiao-bo,TANG Wen-zhi.Alignment evaluation of aspartate aminotransferase detected results between two full automatic biochemical analyzers[J].Laboratory Medicine and Clinic,2014,0(4):455-456.
Authors:XUJian-feng  ZHONG Zhi-juan  GUO Guo-wei  DAI Xiao-bo  TANG Wen-zhi
Institution:1. Zhuhai Hospital of Guangdong Province Traditional Chinese Medical Hospital, Zhuhai, Guangdong 519015, China 2. the Fifth Hos- pital Affiliated to Zhongshan University, Zhuhai, Guangdong 519000, China)
Abstract:Objective To perform alignment evaluation of aspartate aminotransferase (AST)detected results between Hitachi 7170 and Backman DXC800 full automatic biochemical analyzers.Methods Hitachi 7170 analyze was taken as comparison instrument and Backman DXC800 analyzer was taken as test instrument.Experiment was designed according to EP9-A2 document of Clinic and Laboratory Standards Institute (CLSI),fresh serum fasting samples were enrolled.The half value of allowed total error of Clinical Laboratory Improvement Amendments of 1988 (CLIA′88)was taken as the standard to assess the deviation of the detected results of the two analyzers.Results 〈br〉 There was fine correlation of AST detected results between the two analyzers,and the deviation was in clinical ac-ceptable range.Conclusion Serum AST determination results in these two biochemical analyzer might be compara-ble,with fine consistence.
Keywords:alignment evaluation  aspartate aminotransferase  Hitachi 7170 full automatic biochemical an-alyzers  Backman DXC800 full automatic biochemical analyzers
本文献已被 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号