首页 | 本学科首页   官方微博 | 高级检索  
     


Primary activity standardization of Co by sum-peak method
Authors:A. Iwahara   R. Poledna   C.J. da Silva  L. Tauhata
Affiliation:aLaboratório Nacional de Metrologia das Radiações Ionizantes (LNMRI)/Instituto de Radioproteção e Dosimetria (IRD)/Comissão Nacional de Energia Nuclear (CNEN), Av. Salvador Allende, s/no. Recreio dos Bandeirantes – CEP 22780-160 Rio de Janeiro, Brazil
Abstract:The sum-peak method was applied to standardize a 57Co solution within the framework of an international comparison organized by International Atomic Energy Agency, in 2008, aimed toward international traceability of activity measurements. A planar germanium detector was used with the sources placed on top of the detector for activity determination measurements. An analytical expression for accidental summing correction was derived and the effect of the germanium characteristic KX-ray escape peak of 112 keV was taken into account. The standard uncertainty associated to the activity concentration value was 0.37% and the result was compared with other measurement methods.
Keywords:57Co solution   Activity   Sum-peak method   Primary standardization
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏