引用本文:陈刚,吴东,曾蒙苏,张来,顾君英,陆海峰.双源CT冠状动脉成像辐射剂量的优化[J].放射学实践,2014,0(6):593-596.
作者姓名:陈刚  吴东  曾蒙苏  张来  顾君英  陆海峰
作者单位:200032上海,复旦大学附属中山医院放射科;上海市影像医学研究所,复旦大学上海医学院影像学系
摘    要:目的:探讨使用更窄的曝光时间窗对双源CT冠状动脉成像图像质量和辐射剂量的影响。方法:选择心律稳定、配合屏气在本院行冠状动脉CTA检查者共360例,按心率及扫描方案随机分为5组:当心率<65次/分时,行回顾性心电门控扫描,用70%-80%曝光时间窗成像冠状动脉(A组);窄窗方案根据前瞻性心电门控平扫所获得的最佳时相,选择此时相为中心总长1%作为曝光时间窗,行前瞻心电性门控冠状动脉成像(B组)。当心率≥65次/时,行回顾性心电门控扫描,常规使用30%-80%曝光时间窗成像冠状动脉(C组);窄窗方案根据前瞻性心电门控平扫所获得的最佳时相,以此时相为中心选择总长10%的曝光时间窗,分别用前瞻性心电门控(D 组)和低剂量回顾性心电门控行冠状动脉成像(E组);分别计算各组平均有效辐射剂量以及冠状动脉图像质量(4分制评分),并作统计学分析。结果:A 组(8.11±1.54)mSv]和B组(3.63±0.82)mSv],C 组(10.56±2.51)mSv]和 D 组(4.42±1.16)mSv],D 组(4.42±1.16)mSv]和E组(5.43±1.14)mSv]的辐射剂量相比较均有统计学差异(P<0.05),采用窄窗方案前瞻性心电门控技术的辐射剂量最少;而各组间图像质量,均无统计学差异(P>0.05)。结论:前瞻性心电门控通过平扫获得最佳时相,当心率<65次/分时,选择曝光总长1%的时间窗,而当心率≥65次/分时,选择曝光总长10%的时间窗,能获得高质量的、满足诊断要求的图像,并可有效地降低辐射量。

关 键 词:冠状动脉疾病  体层摄影术  X线计算机  血管造影术  辐射量

An optimized radiation dose on dual-source CT coronary angiography
Institution:CHEN Gang, WU Dong, ZENG Meng-su, et al.( Department of Radiology,Zhongshan Hospital, Fudan University, Shanghai Institute of Medical Imaging, Departmentof Medical Imaging, Shanghai Medical College, Fudan University, Shanghai 200032, P. R. China)
Abstract:Objective:To evaluate the image quality and radiation dose of dual-source CT coronary angiography (CT-CA)with a more narrow exposure window.Methods:In all,360 cases underwent dual-source CTCA and were randomly divided into five groups according to the heart rate and scanning protocol.Group A,retrospective ECG-gated scan of the patients with heart rate 〈65 beats per minute (bpm)was performed from 70% to 80%of the R-R interval to reconstruct coronary artery;Group B,prospective ECG-gated scan of the patients with heart rate 〈65bpm was performed with 1% of the total length of the exposure window;Group C,retrospective ECG-gated scan of the patients with heart rate ≥65bpm was performed from 30% to 80% of the R-R interval to reconstruct coronary artery;Group D,prospective ECG-gated scan of the patients with heart rate ≥65bpm was performed with 10% exposure window;Group E,low dose retrospective ECG-gated scan of the patients with heart rate ≥65bpm was performed.The average effective radiation dose and image quality of coronary arteries (4-point scale)were calculated for each group,and statistical analysis was done.Results:Difference of radiation dose between each group was statistically significant (P〈0.05).A narrowed scan window on prospective ECG gated dual-source CTCA could reduce the radiation dose;while there was no significant difference in the image quality between the groups (P〉0.05).Conclusion:Prospective ECG-gated dual-source CTCA can reduce the radiation dose while maintaining good image quality by selecting 1% of the total length of the exposure window(heart rate 〈65bpm),and 10% of the total length of the exposure window (heart rate ≥65bpm)after obtaining the optimal interval by plain scan.
Keywords:Coronary artery disease  Tomography  X-ray computed  Angiography  Radiation dosage