首页 | 本学科首页   官方微博 | 高级检索  
检索        


Dural Tissue Trauma and Cerebrospinal Fluid Leak after Epidural Needle Puncture: Effect of Needle Design, Angle, and Bevel Orientation
Authors:Angle  Pamela J MD  FRCPC; Kronberg  Jean E MD  PhD  FRCPC&#x;; Thompson  Dorothy E MB  FRCPC; Ackerley  Cameron PhD&#x;; Szalai  John Paul PhD ; Duffin  James PhD&#x;; Faure  Peter RN#
Institution:Angle, Pamela J. M.D., F.R.C.P.C.*; Kronberg, Jean E. M.D., Ph.D., F.R.C.P.C.†; Thompson, Dorothy E. M.B., F.R.C.P.C.*; Ackerley, Cameron Ph.D.‡; Szalai, John Paul Ph.D.§; Duffin, James Ph.D.∥; Faure, Peter R.N.#
Abstract:Background: The effects of epidural needle design, angle, and bevel orientation on cerebrospinal fluid leak after puncture have not been reported. The impact of these factors on leak rate was examined using a dural sac model. Dural trauma was examined using scanning electron microscopy.

Methods: Human cadaveric dura, mounted on a cylindrical model, was punctured with epidural needles using a micromanipulator. Tissue was punctured at 15 cm H2O (left lateral decubitus) system pressure, and leak was measured at 25 cm H2O (semisitting) pressure. Leak rates and trauma were compared for the following: (1) six different epidural needles at 90degrees], bevel parallel to the dural long axis; (2) 18-gauge Tuohy and 18-gauge Special Sprotte(R) epidural needles, 30degrees]versus 90degrees]; (3) 18-gauge Tuohy, bevel perpendicular versus parallel to the dural long axis.

Results: With the 90degrees] puncture, bevel parallel, the greatest leak occurred with a 17-gauge Hustead (516 +/- 319 ml/15 min), and the smallest leak occurred with a 20-gauge Tuohy (100 +/- 112 ml/15 min; P = 0.0018). A 20-gauge Tuohy puncture led to statistically significant reductions in leak (P value range, 0.0001-0.0024) compared with all needles except the Special Sprotte(R). With the 30degrees]versus 90degrees] angle, 30degrees] punctures with an 18-gauge Tuohy produced nonstatistically significant leak reductions compared with the 18-gauge Tuohy at 90degrees]. The puncture angle made no difference for the Special Sprotte(R). Nonsignificant reductions were found for the Special Sprotte(R) compared with the Tuohy. With the 18-gauge Tuohy bevel orientation, perpendicular orientation produced nonstatistically significant reductions in leak compared with parallel orientation.

Keywords:
点击此处可从《The Journal of the American Society of Anesthesiologists》浏览原始摘要信息
设为首页 | 免责声明 | 关于勤云 | 加入收藏