In vivo study of an elastomer end-coated polytetrafluoroethylene vascular prosthesis. |
| |
Authors: | H Tsuchida S E Wilson R Kowligi S K Peng |
| |
Affiliation: | Department of Surgery, Harbor-UCLA Medical Center, Torrance 90509. |
| |
Abstract: | Polyurethane end-coated polytetrafluoroethylene (PTFE) grafts (elastomer PTFE grafts) were implanted in 12 female adult mongrel dogs to assess patency, intimalization, tissue incorporation, and technical suitability of the material as a vascular graft. Each dog had bilateral aortoiliac grafts placed, one a standard PTFE and the other an elastomer PTFE graft. The length of the grafts was 7-8 cm and the diameter was 6 mm. The grafts were harvested at intervals to 120 days postoperatively. The elastomer PTFE grafts showed superior longitudinal elasticity, retention of shape, and no graft tearing with suture tension; however, no significant difference in bleeding was noted at the anastomoses between the standard and elastomer PTFE grafts. Satisfactory patency was obtained with both standard (8/10) and elastomer PTFE grafts (9/10) at 90-120 days. No significant difference in the thickness of intima and the length of pannus ingrowth was noted between the standard and elastomer PTFE grafts. No outer tissue incorporation was seen at the elastomer-treated graft segments as opposed to the well-incorporated untreated segments. In conclusion, elastomer end-coating of a PTFE vascular prosthesis provided excellent handling characteristics without detracting from patency; however, the lack of outer tissue incorporation may be a potential disadvantage in its clinical use. |
| |