Octavius 4D剂量验证系统应用及误差因素分析 |
| |
引用本文: | 雷伟杰,陆启勇.Octavius 4D剂量验证系统应用及误差因素分析[J].医疗装备,2014(12):1-3. |
| |
作者姓名: | 雷伟杰 陆启勇 |
| |
作者单位: | 徐州矿务集团总医院放疗中心 |
| |
摘 要: | 本文首先阐述了Octavius 4D剂量验证系统原理及应用,对其用于VMAT放疗计划进行剂量验证的实施方法、验证方式等进行了分析,探讨了剂量验证误差的影响因素.
|
关 键 词: | 剂量验证 VMAT放射治疗 |
Application of Octavius 4D Dose Verification System and Analysition of Dose Deviation Factors |
| |
Institution: | LEI Wei - jie, LU Qi - yong (Xuzhou Miner General Hospital, Xuzhou 21006, China) |
| |
Abstract: | The principle was described for the Octavius 4D dose verification system, and the implementation methods of VMAT plan dose verification and verification methods were analyzed. A tree diagram of factors which affected gamma passing rate was set up for cheeking reasons in dose verifieation of VMAT plans. |
| |
Keywords: | Octavius 4D Dose verification VMAT Radiotherapy |
本文献已被 CNKI 维普 等数据库收录! |