首页 | 本学科首页   官方微博 | 高级检索  
     


Development of Digital Subtraction Angiography for Coronary Artery
Authors:Megumi Yamamoto  Yasuhiko Okura  Masaharu Ishihara  Masayuki Kagemoto  Kengo Harada  Takayuki Ishida
Affiliation:(1) Major in Medical Engineering and Technology, Graduate Course in Integrated Human Studies, Hiroshima International University, 555-36 Kurose-Gakuendai, Higashi Hiroshima, 739-2695, Japan;(2) Department of Cardiology, Hiroshima City Hospital, 7-33, Nakaku-Motomachi, Hiroshima 730-8518, Japan;(3) Department of Radiology, Hiroshima City Hospital, 7-33, Nakaku-Motomachi, Hiroshima 730-8518, Japan
Abstract:The purpose of this research is to develop a new method of digital subtraction angiography (DSA) that can be applied to real time with reducing motion artifacts caused by heart movement and respiration. To create the mask image for DSA, the maximum pixel value at each pixel (which is the opposite pixel value to that of a vessel filled by contrast medium) was selected from the previous 14 image frames. The search area for the maximum pixel value was selected using the value of the standard deviation (SD) for each pixel from the previous 14 image frames. When the SD value in the 14 frames was greater than a threshold level, the search area of the maximum value became 1 pixel × 1 pixel × 14 frames; otherwise, 7 pixels × 7 pixels × 7 frames. The image quality of new DSA was evaluated on 20 coronary arteriogram images, including various degrees of occlusion or stenosis. The results indicated a considerable improvement in DSA image quality; thus, the coronary arteries, carotid artery, and vein were clearly enhanced.
Keywords:Digital subtraction angiography (DSA)  motion artifact  misregistration  coronary angiography
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号