网织红细胞镜检法与全自动法比较分析
引用本文:褚静英,羽晓瑜,陆玉霞,徐廷云,孙中文,徐文慧.网织红细胞镜检法与全自动法比较分析[J].检验医学,2010,25(11):841-843.
作者姓名:褚静英  羽晓瑜  陆玉霞  徐廷云  孙中文  徐文慧
作者单位:苏州卫生职业技术学院检验药学系,江苏,苏州,215009
摘    要:目的比较镜检法与全自动法检测网织红细胞(Ret)结果的准确性。方法针对不同部位采血的标本(末梢血或静脉血)和不同前处理的标本分别采用镜检法和全自动法作Ret计数。结果不同处理标本Ret检测结果比较:不抗凝静脉血(0.921±0.402)%,乙二胺四乙酸二钾(EDTA-K2)抗凝静脉血(1.022±0.399)%,两者差异无统计学意义(P〉0.05);不同部位采血的标本Ret检测结果比较:不抗凝静脉血为(1.262±0.519)%,不抗凝末梢血为(1.159±0.481)%,两者差异无统计学意义(P〉0.05);EDTA-K2抗凝静脉血为(1.371±0.399)%,EDTA-K2抗凝末梢血为(1.342±0.500)%,两者差异无统计学意义(P〉0.05);不同方法检测Ret的结果比较:全自动法为(1.326±0.461)%,镜检法为(1.386±0.418)%,两者差异无统计学意义(P〉0.05)。结论对于健康正常人,规范化Ret计数结果不受抗凝剂的使用、不同部位采血及检测方法的影响。

关 键 词:网织红细胞  镜检法  全自动法  静脉血  末梢血

Comparison analysis of microscope test and automatic test of reticulocyte count
CHU Jingying,YU Xiaoyu,LU Yuxia,XU Tingyun,SUN Zhongwen,XU Wenhui.Comparison analysis of microscope test and automatic test of reticulocyte count[J].Laboratory Medicine,2010,25(11):841-843.
Authors:CHU Jingying  YU Xiaoyu  LU Yuxia  XU Tingyun  SUN Zhongwen  XU Wenhui
Institution:.(Faculty of Laboratory Medicine and Pharmacy,Suzhou Health College,Jiangsu Suzhou 215009,China)
Abstract:Objective To compare the accuracy of microscope test and automatic test of reticulocyte(Ret) count.Methods The microscope test and automatic test were used to count Ret,and the samples were collected from different resources(capillary blood or venous blood)or were treated differently before being tested.Results As the sample being treated differently,the values of Ret count were(0.921±0.402)% in the venous blood without anticoagulant and(1.022±0.399)% in the venous blood with dipotassium ethylenediamine tetraacetic acid(EDTA-K2),and there was no significant difference in Ret count between anticoagulant blood and no anticoagulant blood(P0.05).As the sample from different resources,the values of Ret count were(1.262±0.519)% in the venous blood without anticoagulant,(1.159±0.481)% in the capillary blood without anticoagulant,(1.371±0.399)% in the venous blood with EDTA-K2 and(1.342±0.500)% in the capillary blood with EDTA-K2,and there was no significant difference in Ret count between venous blood and capillary blood(P0.05).As using different methods to test the samples,the Ret count results were(1.326±0.461)% by automatic test and(1.386±0.418)% by microscope test,and there was no significant difference in Ret count between microscope test and automatic test(P0.05).Conclusions For healthy people,the anticoagulant,sample resource and determination method will not influence on the Ret count result.
Keywords:Reticulocyte  Microscope test  Automatic test  Venous blood  Capillary blood