首页 | 本学科首页   官方微博 | 高级检索  
     

血清胰岛素TRFIA的建立
引用本文:姚焱,李国祥,陈键. 血清胰岛素TRFIA的建立[J]. 标记免疫分析与临床, 2012, 19(1): 36-39
作者姓名:姚焱  李国祥  陈键
作者单位:原子高科股份有限公司,北京,102413;中国同位素有限公司,北京,100045
摘    要:目的 建立血清胰岛素( Ins)时间分辨荧光免疫分析法(time-resolved fluoroimmunoassay,TRFIA).方法 采用一株Ins-MAb包被于固相微孔板上,另一株Ins-MAb用DTTA- Eu3+标记作为示踪剂采用Victor 1420荧光仪测定,.并对本法进行评价.结果 本法的标准曲线范围...

关 键 词:胰岛素    时间分辨荧光免疫分析  单克隆抗体  血清

Establishment of TRFIA for Serum Insulin
YAO Yan , LI Guo-xiang , CHEN Jian. Establishment of TRFIA for Serum Insulin[J]. Labeled Immunoassays and Clinical Medicine, 2012, 19(1): 36-39
Authors:YAO Yan    LI Guo-xiang    CHEN Jian
Affiliation:(Atom High Tech Co., Ltd, Beijing 102413, China)
Abstract:Objective A time-resolved fluoroimmunoassay (TRFIA) for the measurement of insulin (Ins) in human serum had been established. Methods The test was based on the one-step sandwich technique with two MAbs directed against two different epitopes of Ins using coated 96-well mierotitre plate as the solid phase and DITA-Eu3+ as the label, and the established method was evaluated with Victor 1420 muhilabel counter. Results The standard ranges were 5 - 160mU/L. The detection limit was 0.35mU/L. At three different levels of Ins, the intra-assay CV was 0.90% to 4.10%, the inter-assay CV was 4.55% to 4.96%. The average re- covery rates were 97.7% - 102.9%. The detected results with this method closely corresponded to those of by ELISA with DSL (R = 0. 982, n = 80). Conclusion The established TRFIA in this study was specific, sensi- tive, easy to use, and suitable for clinical serum insulin measurement.
Keywords:Insulin  Europium  Time-resolved fluoroimmunoassay  Monoantibody  Serum