首页 | 本学科首页   官方微博 | 高级检索  
     


Periprosthetic Fracture Torque for Short Versus Standard Cemented Hip Stems: An Experimental In Vitro Study
Authors:Takkan MorishimaBastiaan L. Ginsel  BHB  MbChB  Godwin G.H. ChoyLance J. Wilson  BE Mech  ME Sc  PhD  Sarah L. WhitehouseRoss W. Crawford  MBBS  DPhil Oxon
Affiliation:Institute of Health and Biomedical Innovation, Queensland University of Technology, The Prince Charles Hospital, Brisbane, Queensland, Australia
Abstract:In an attempt to preserve proximal femoral bone stock and achieve a better fit in smaller femora, especially in the Asian population, several new shorter stem designs have become available. We investigated the torque to periprosthetic femoral fracture of the Exeter short stem compared with the conventional length Exeter stem in a Sawbone model. Forty-two stems; 21 shorter and 21 conventional stems both with three different offsets were cemented in a composite Sawbone model and torqued to fracture. Results showed that Sawbone femurs break at a statistically significantly lower torque to failure with a shorter compared to conventional-length Exeter stem of the same offset. Both standard and short-stem designs are safe to use as the torque to failure is 7–10 times that seen in activities of daily living.
Keywords:short femoral stem   fracture load   stress shielding   biomechanical study   sawbones
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号