首页 | 本学科首页   官方微博 | 高级检索  
     


Preliminary evaluation of second harmonic direct detection scheme for low-dose range in alanine/EPR dosimetry
Authors:Chen Felipe  Graeff Carlos F O  Baffa Oswaldo
Affiliation:Departamento de Física e Matemática, FFCLRP, Universidade de S?o Paulo, Brazil.
Abstract:The usefulness of a direct detection scheme of the second harmonic (2h) overmodulated signal from irradiated alanine in EPR dosimetry was studied. For this purpose, a group of DL-alanine/paraffin cylindrical pellets was produced. The dosimeters were irradiated with a 60Co radiotherapy gamma source with doses of 0.05, 0.1, 0.5, 1 and 5 Gy. The EPR measurements were carried out in a VARIAN-E4 spectrometer operating in X-band with optimized parameters to obtain highest amplitude signals of both harmonics. The 2h signal was detected directly at twice the modulation frequency. In preliminary results, the 2h showed some advantages over the 1 h such as better resolution for doses below 1 Gy, better repeatability results and better linear behaviour in the dose range indicated.
Keywords:
本文献已被 PubMed 等数据库收录!