两次激发SE-EPI在肝脏MRI的应用
引用本文:丁荏芬,郭友,陈曌,郑晓林.两次激发SE-EPI在肝脏MRI的应用[J].影像诊断与介入放射学,2005,14(2):101-104.
作者姓名:丁荏芬  郭友  陈曌  郑晓林
作者单位:523018,广东东莞市人民医院放射科
摘    要:目的比较两次激发SE-EPI与呼吸门控FSE及SSFSET2WI在肝脏的应用。方法对14名志愿者及21名肝病患者行上腹部呼吸门控FSE及SSFSE和屏气2次激发SE-EPI(TR=5999ms和1999ms)扫描。所有T2WI序列均运用脂肪抑制技术。定量分析肝脏,病灶的信噪比及肝脏-病灶的对比度噪声比,评价各序列的图像质量及伪影。结果两次激发SE-EPI与SSFSE及FSE在肝脏及病灶信噪比,肝脏-病灶对比度噪声比和图像质量方面无明显差别(P>0.05)。SE-EPI的磁敏感伪影及化学位移伪影均较FSE及SSFSE重,其运动伪影较FSE轻(P<0.05)。结论两次激发SE-EPI能够在较短时间里提供较高质量的上腹部T2WI。被检查者在扫描时可自由平静呼吸或屏气,可作为肝脏T2WI选择序列之一。

关 键 词:肝脏  磁共振成像  平面回波成像
修稿时间:2004年10月15

Two-shot SE-EPI in the hepatic MR imaging
DING Renfen,GUO You,CHEN Zhao,et al..Two-shot SE-EPI in the hepatic MR imaging[J].Journal of Diagnostic Imaging & Interventional Radiology,2005,14(2):101-104.
Authors:DING Renfen  GUO You  CHEN Zhao  
Institution:DING Renfen,GUO You,CHEN Zhao,et al. Department of Radiology,Dongguan People's Hospital,Dongguan 523018
Abstract:Objective To compare the performance of T2 -weighted two-shot sp in-echo version of echo planar imaging (SE-EPI), respiratory-gating fast spin -echo (FSE) and single-shot fast spin echo (SSFSE) sequences for the evaluatio n of focal hepatic lesions. Methods Breath-hold two-shot SE-EPI (TR=5999ms and 1999ms) of the liver was performed in 14 healthy volunteers and 21 patients, a nd was compared with respiratory-gating T2-weithted FSE and SSFSE. The fat-su ppression technique was used in each sequence. The signal-to-noise ratio (SNR) of the liver and lesion and the contrast-to-noise ratio (CNR) of the l esion-to-liver were quantitatively analyzed. The image quality and artifacts w ere also comparatively studied. Results Quantitatively, SE_EPI, SSFSE and FSE se quences showed non-significant differences (P > 0.05) in liver/lesion CNR, SNR of Liver and lesion and the general image quality. Susceptibility artifacts and chemical shift artifacts on SE-EPI were more than SSFSE and FSE. There were few er motion and ghost artifacts on SE-EPI images compared to FSE images (P< 0.05) . Conclusion SE-EPI of the liver can provide better quality T2-weighted images in shorter acquisition times compared with respiratory-gating FSE and SSFSE. There is no particular requirement for patients to suspend their breath during the aquisition. This technique may be useful in clinical application of hepatic MR imaging.
Keywords:Liver  Magnetic resonance imaging  Echo planar imaging (EPI)