成人胸椎椎弓根横径对螺钉植入安全性的影响
引用本文:付长峰,刘中国,刘一,张绍昆,赵松,闫明.成人胸椎椎弓根横径对螺钉植入安全性的影响[J].吉林大学学报(医学版),2009,35(2):246-248.
作者姓名:付长峰  刘中国  刘一  张绍昆  赵松  闫明
作者单位:(1.吉林大学第一医院脊柱外科,吉林 长春 130021;2.福建省厦门市第三医院骨科, 福建 厦门361100 )
基金项目:国家自然科学基金,吉林省卫生厅重点科室基金 
摘    要:目的:评价成人上位、中位胸椎节段(T1~T8)椎弓根形态对螺钉植入安全性的影响。方法:采用大体直接测量方法,测量5具新鲜成人尸体胸椎(T1~T8)椎弓根的横径,比较其与临床应用的胸椎椎弓根螺钉的匹配程度;模拟手术情况,按照个体化原则选择植入螺钉型号,并于直视下植入椎弓根螺钉,肉眼下观察螺钉植入后椎弓根皮质破坏情况。结果:成人胸椎T6~T8节段,椎弓根横径分别为4.53(4.40~4.70)、5.38(5.25~5.60)和5.55(5.50~5.80)mm,均>4.5 mm,可以满足最小直径螺钉植入。T1~T5椎弓根横径分别为4.74(3.85~5.25)、4.47(3.45~5.25)、4.15(3.60~4.65)、3.75(3.10~4.50)和4.25(3.70~4.65)mm。在所有测量胸椎标本中,椎弓根横径<4.5 mm的椎体占标本总体的22.5%。螺钉植入椎弓根后,2个椎体发生椎弓根外侧皮质破坏。结论:成人T3~T4节段椎弓根横径最窄,影响椎弓根螺钉的安全植入,但在上位及中位胸椎的大部分节段,可以安全植入。

关 键 词:椎弓根  安全性    
收稿时间:2008-09-26

Influence of pedicle transverse diameter on security of screw implantation in adult thoracic vertebrae
FU Chang-feng,LIU Zhong-guo,LIU Yi,ZHANG Shao-kun,ZHAO Song,YAN Ming.Influence of pedicle transverse diameter on security of screw implantation in adult thoracic vertebrae[J].Journal of Jilin University: Med Ed,2009,35(2):246-248.
Authors:FU Chang-feng  LIU Zhong-guo  LIU Yi  ZHANG Shao-kun  ZHAO Song  YAN Ming
Institution:(1.Deartpment of Orthopaedics,First Hospital,Jilin University,Changchun 130021,China;2.Deartpment of Orthopaedics,Third Hospital of Xiamen City,iamen 361100,China)
Abstract:Objective To identify the effect of morphology of adult upper and middle thoracic pedicle on security of screw implantation.Methods Five fresh adult cadaveric thoracic spine from T1 to T8 were harvested.Pedicle transverse diameters were measured and compared with the smallest diameter(4.5 mm) of screw used in clinic.The screws were inserted and implant effect was observed.Results The mean transverse diameter of adult thoracic(T1-T8) were T1:4.74(3.85-5.25)mm,T2:4.51(3.45-5.25)mm,T3:4.15(3.60-4.65)mm,T4:3.75(3.10-4.50)mm,T5:4.25(3.70-4.65)mm,T6: 4.53(4.40-4.70)mm,T7:5.38(5.25-5.60)mm,T8:5.55(5.50-5.80)mm.Of all specimens,there were 9 specmen(22.5%) whose tranverse diameters were smaller than 4.5 mm.After implantation,two screws broke through pedicle lateral wall.Conclusion Adult pedicle transverse diameters in T3-T4 are smallest,transpedicular screw placement is difficult in some of these vertebrae,but in most of thoracic vertebrae,pedicle tranverse diameter could permit screw implanting in security.
Keywords:thoracic vertebrae  pedicle  security