Effect of exposure time on curing efficiency of polymerizing units equipped with light-emitting diodes
Authors:David José R  Gomes Osnara M  Gomes João C  Loguercio Alessandro D  Reis Alessandra
Institution:Department of Dentistry, School of Dentistry, University of Ponta Grossa, Campus Uvaranas, Brazil.
Abstract:A study was conducted to evaluate the top and bottom hardness of two composites cured using polymerizing units equipped with light-emitting diodes LED] (LEDemetron; Elipar FreeLight, Coltolux LED) and one quartz-tungsten halogen device QTH] (Optilux 501) under different exposure times (20, 40 and 60 sec). A matrix mold 5 mm in diameter and 2 mm in depth was made to obtain five disc-shaped specimens for each experimental group. The specimens were cured by one of the light-curing units (LCUs) for 20, 40 or 60 sec, and the hardness was measured with a Vickers hardness-measuring instrument (50 g/30 sec). Data were subjected to three-way ANOVA and Tukey's test (alpha = 0.05). LED LCUs were as effective as the QTH device for curing both composites. A significant increase in the microhardness values were observed for all light LCUs when the exposure time was changed from 20 sec to 40 sec. The Z250 composite showed hardness values that were usually higher than those of the Charisma composite under similar experimental conditions. LED LCUs are as efficient for curing composites as the QTH device as long as an exposure time of 40 sec or higher is employed. An exposure time of 40 sec is required to provide composites with a uniform and high Knoop hardness when LED light-curing units are employed.