Quality assurance program for whole blood prothrombin time-international normalized ratio point-of-care monitors used for patient self-testing to control oral anticoagulation
Authors:Tripodi Armando  Bressi Caterina  Carpenedo Monica  Chantarangkul Veena  Clerici Marigrazia  Mannucci Pier Mannuccio
Affiliation:Angelo Bianchi Bonomi Hemophilia and Thrombosis Center, Department of Internal Medicine, University and IRCCS Maggiore Hospital, Via Pace 9, 20122 Milan, Italy
Abstract:Whole blood coagulation monitors are increasingly used for patient self-testing to control oral anticoagulation, but there are no comprehensive quality assurance (QA) programs to check their performance. We report on the experience with one of such programs applied in a field study where patients on prothrombin time (PT)-international normalized ratio (INR) self-testing were asked to bring their monitors to the anticoagulation clinic for checking.PT-INR testing was performed three times over 3 months with 14 patient's monitors and test strips on three recalcified QA plasmas by an experienced laboratory operator. Each patient was also asked to perform PT-INR self-testing (his/her own capillary blood) which was then compared to the laboratory PT-INR (plasma).Overall, the comparison between the observed and the consensus PT-INR on QA plasmas was acceptable with the majority of measurements lying within ±15% or 20% of the consensus values. The comparison between the PT-INR self-testing and the laboratory method was also acceptable: overall, there was no statistical significant difference between the mean PT-INR values and the majority of paired measurements were less than 15% or 20% apart.In conclusion, our results show that the proposed QA scheme is feasible and may be implemented on a larger scale. Monitors should be recalled periodically to the clinic where they have been prescribed to the patient. During each visit, the clinic may check the monitors and patient self-testing performance as described. Such comprehensive QA system would make monitoring of oral anticoagulant treatment by self-testing safer and more effective.
Keywords:Oral anticoagulant treatment   INR   Self-testing, point of care   Monitor
本文献已被 ScienceDirect PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号