首页 | 本学科首页   官方微博 | 高级检索  
     


High-energy radiation monitoring based on radio-fluorogenic co-polymerization II: fixed fluorescent images of collimated x-ray beams using an RFCP gel
Authors:Warman J M  Luthjens L H  de Haas M P
Affiliation:Faculty of Applied Sciences, Department of Radiation and Isotopes for Health, Delft University of Technology, Delft, The Netherlands. J.M.Warman@TUDelft.NL
Abstract:We have produced an optically clear, close to water-equivalent gel that is radio-fluorogenic, i.e. fluoresces in UV light after exposure to high-energy radiation. Its potential as a 2D and 3D dosimetric medium is demonstrated by fixed fluorescent images of the cross-section, track and intersection of collimated (10 × 10 or 5 × 5 mm(2)) 205 kVp x-ray beams. The images, produced by doses on the order of 10 Gy, are formed instantaneously and can be digitally recorded and scanned with a spatial resolution on the order of 0.1 mm. No loss of spatial resolution occurs on standing under ambient conditions for at least 3 days.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏