High-field MRS of the human brain at short TE and TR
Authors:Boer Vincent O  Siero Jeroen C W  Hoogduin Hans  van Gorp Jetse S  Luijten Peter R  Klomp Dennis W J
Affiliation:Department of Radiology, Radiotherapy and Nuclear Medicine, UMC Utrecht, Utrecht, The Netherlands. V.O.Boer@umcutrecht.nl
Abstract:In vivo MRS of the human brain at 7 tesla allows identification of a large number of metabolites at higher spatial resolutions than currently possible at lower field strengths. However, several challenges complicate in vivo localization and artifact suppression in MRS at high spatial resolution within a clinically feasible scan time at 7 tesla. Published MRS sequences at 7 tesla suffer from long echo times, inherent signal‐to‐noise ratio (SNR) loss, large chemical shift displacement artifacts or long repetition times because of excessive radiofrequency (RF) power deposition. In the present study a pulse‐acquire sequence was used that does not suffer from these high field drawbacks. A slice selective excitation combined with high resolution chemical shift imaging for in‐plane localization was used to limit chemical shift displacement artifacts. The pulse‐acquire approach resulted in a very short echo time of 1.4 ms. A cost function guided shimming algorithm was developed to constrain frequency offsets in the excited slice, therefore adiabatic frequency selective suppression could be employed to minimize artifacts from high intensity lipids and water signals in the excited slice. The high sensitivity at a TR of 1 s was demonstrated both on a supraventricular slice as well as in an area very close to the skull in the frontal cortex at a nominal spatial resolution of 0.25 cc within a feasible scan time. Copyright © 2011 John Wiley & Sons, Ltd.
Keywords:human brain  7T  proton MRS  high resolution  short echo time  FID
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号