首页 | 本学科首页   官方微博 | 高级检索  
检索        

金宝血透机AK200维修案例
引用本文:王上东,王威.金宝血透机AK200维修案例[J].中国医学装备,2009,6(10):50-51.
作者姓名:王上东  王威
作者单位:汕头市中心医院,广东,汕头,515031
摘    要:金宝AK200自检能正常通过,但干粉桶进水不正常,然后电导报警。这是血透机使用过程经常会碰到的故障。文章根据图解详细分析了血透机的工作原理,根据故障现象准确查出故障原因所在,对同类设备的维修具有一定的指导意义。

关 键 词:电导  干粉桶  O形圈  滤网  

Troubleshooting for GAMBRO AK200 hemodialysis machine
Wang Shangdong,Wang Wei.Troubleshooting for GAMBRO AK200 hemodialysis machine[J].China Medical Equipment,2009,6(10):50-51.
Authors:Wang Shangdong  Wang Wei
Institution:(Shantou Central Hospital, Shantou 515031, China.)
Abstract:The GAMBRO AK200 warning about the conductivity after selftest because water can't prime into the BI-CART. We'll find this wrong component frequently. This article analyzed the operation principle of this equipment in detail based on its electric circuit, and located the wrong component rapidly according its trouble phenomenon. We hope this could be a help in the maintenance of the same kind of equipments.
Keywords:Conductivity  Bi-Cart  O-ring  Filter  Valve
本文献已被 维普 万方数据 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号