牦牛心包瓣置换二尖瓣术后瓣膜毁损
引用本文:肖锡俊 石应康. 牦牛心包瓣置换二尖瓣术后瓣膜毁损[J]. 中国胸心血管外科临床杂志, 1994, 1(1): 32-35
作者姓名:肖锡俊 石应康
作者单位:华西医科大学附属一院胸心外科,华西医科大学附属一院超声科
摘    要:1985至1987年用牦牛心包瓣置换二尖瓣。术后65例随访至1993年9月,累积随访379.4病人年,平均6.9年。5年存活率75%。晚期死亡9例,其中置换瓣膜心内膜炎(PVE)5例,血栓栓塞(TE)、瓣膜毁损、心衰、原因不明各1例,死亡率3.69%病人年。PVE、TE发生率分别为1.32%,0.26%病人年。瓣膜毁损指:(1)瓣膜钙化、穿孔或撕裂,或(2)瓣膜中度以上狭窄或返流。瓣膜毁损可进一步分为轻度和重度。随访期间瓣膜毁损16例,瓣膜毁损率4.22%病人年。作者认为:牦牛心包瓣置换二尖瓣术后远期结果是可接受的,但术后5年开始应加强随访。为便于临床治疗及评价,生物瓣毁损可进一步分为轻度和重度。

关 键 词:牦牛心包瓣,二尖瓣置换,瓣膜毁损

Structural Valve Deterioration after Mitral Valve Replacement with Yak Pericardial Bioprostheses
Xiao Xi-jun, Shi Ying-kang,Tian Zi-pu,et al.. Structural Valve Deterioration after Mitral Valve Replacement with Yak Pericardial Bioprostheses[J]. Chinese Journal of Clinical Thoracic and Cardiovascular Surgery, 1994, 1(1): 32-35
Authors:Xiao Xi-jun   Shi Ying-kang  Tian Zi-pu  et al.
Affiliation:Xiao Xi-jun, Shi Ying-kang,Tian Zi-pu, et al.
Abstract:From l986 to 1987, Yak pericardial Bioprostheses were inSerted as the mitral valve replacement substitute. After operation , 55 cases were followed to September 1993. The follow-up totalled 379. 4 patient-years,mean 6.9 years.5-year actural survial rate was 75%. There were 9 cases of deaths.The deaths were associated with PVE,TE,heart failure and structural valve deterioration except one. The late mortality rate was 3.69% patient years. PVE, and TE were 1.32%, 0.26% patient-years respectively.Structural valve deterioration is defined as: (1)bioprosthetic valve calclfication , perforation, or disruption; Dr(2) moderate or severe bioprosthetic valve stenosis or regurgitation, The deterioration Was further divided into the mild and the severe, There were 16 cases of structutal valve deterioration.The structural valve deterioration rate was 4.22% patient-years,It is believed that the long-term results of the mitral valve replacement with Yak Pericardial Bioprtatheses are acceptable. Because of incReasing cases of structural valved eterioration. there patients should be followed up carefully from the fifth year after operation. In order to treat and evaluate the patients, the structural valve deterioration of bioprostheses should be further divided into the mild and the severe.
Keywords:Yak pericardial bioprostheses  Mitral valve replace moent  Structural valve deterioration