Inappropriate Shock Delivery Due to Interference between a Washing Machine and an Implantable Cardioverter Defibrillator
Authors:Christof Kolb  Sebastian Schmieder  Claus Schmitt
Affiliation:(1) Deutsches Herzzentrum und 1. Med. Klinik, Klinikum rechts der Isar der TU München, Munich, Germany
Abstract:Electromagnetic interference with implantable cardioverter defibrillators (ICD) can cause inappropriate delivery of therapies or temporary inhibition of ICD functions. The presented case describes electromagnetic interference between a washing machine and an ICD resulting in an inappropriate discharge of the device due to false detection of ventricular fibrillation.
Keywords:electromagnetic interference  inappropriate therapy  ICD