The use of Teflon-embedded calcium sulphate: dysprosium thermoluminescent dosimeters for chest radiography
Authors:C Borrás  M M Liss  C Thomason  O Hidalgo-Salvatierra  C P White  J E Barton
Institution:Western Center for Radiological Physics, West Coast Cancer Foundation, San Francisco, CA 94133.
Abstract:As part of a U.S. multi-regional pilot study conducted by the six Centers for Radiological Physics, 12-mm-diameter by 0.4-mm-thick CaSO4:Dy Teflon-embedded discs were evaluated and used to measure patient entrance exposure on 60 "average" patients at 12 clinical centers. The discs were found to have adequate sensitivity, reproducibility and linearity up to 69.7 microC kg-1 (270 mR). The minimum measurable exposure was estimated as 0.4 microC kg-1 (1.5 mR). All responses were corrected for energy dependence, which varied +/- 20% from 1.7 to 6.5 mm Al half-value layer. Patient entrance exposure values ranged from 1.3 to 28 microC kg-1 (5 to 110 mR), with a median value of 5.2 microC kg-1 (20 mR). This value agreed with exposure measurements made on the chest radiography equipment using an ionization chamber and a phantom which simulated an "average" patient, and with published Nationwide Evaluation of X-Ray Trends (NEXT) data for the same period.