Band artifacts due to bulk motion.
Authors:Pippa Storey  Qun Chen  Wei Li  Robert R Edelman  Pottumarthi V Prasad
Institution:Department of Radiology, MRI Research, Evanston Hospital, vanston Northwestern Healthcare, Evanston, Illinois 60201, USA. p-storey@northwestern.edu
Abstract:Band artifacts due to bulk motion were investigated in images acquired with fast gradient echo sequences. A simple analytical calculation shows that the width of the artifacts has a square-root dependence on the velocity of the imaged object, the time taken to acquire each line of k-space and the field of view in the phase-encoding direction. The theory furthermore predicts that the artifact width can be reduced using parallel imaging by a factor equal to the square root of the acceleration parameter. The analysis and results are presented for motion in the phase- and frequency-encoding directions and comparisons are made between sequential and centric ordering. The theory is validated in phantom experiments, in which bulk motion is simulated in a controlled and reproducible manner by rocking the scan table back and forth along the bore axis. Preliminary cardiac studies in healthy human volunteers show that dark bands may be observed in the endocardium in images acquired with nonsegmented fast gradient echo sequences. The fact that the position of the bands changes with the phase-encoding direction suggests that they may be artifacts due to motion of the heart walls during the image acquisition period.
Keywords:MRI  motion artifacts  parallel imaging  phase‐encoding order  myocardium