Accuracy of Real-Time Single- and Multi-Beat 3-D Speckle Tracking Echocardiography In Vitro
Authors:Johannes Just Hjertaas,Henrik Fosså  ,Grete Lunestad Dybdahl,Renate Grü  ner,Per Lunde,Knut Matre
Affiliation: Institute of Medicine, University of Bergen, Bergen, Norway; Department of Physics and Technology, University of Bergen, Bergen, Norway
Abstract:With little data published on the accuracy of cardiac 3-D strain measurements, we investigated the agreement between 3-D echocardiography and sonomicrometry in an in vitro model with a polyvinyl alcohol phantom. A cardiac scanner with a 3-D probe was used to acquire recordings at 15 different stroke volumes at a heart rate of 60 beats/min, and eight different stroke volumes at a heart rate of 120 beats/min. Sonomicrometry was used as a reference, monitoring longitudinal, circumferential and radial lengths. Both single- and multi-beat acquisitions were recorded. Strain values were compared with sonomicrometer strain using linear correlation coefficients and Bland–Altman analysis. Multi-beat acquisition showed good agreement, whereas real-time images showed less agreement. The best correlation was obtained for a heart rate 60 of beats/min at a volume rate 36.6 volumes/s.
Keywords:In   vitro   Phantom   Left ventricle   Echocardiography   Three-dimensional ultrasound   Strain   Sonomicrometer   Volume rate   Frame rate   Polyvinyl alcohol