Effect of implant connection and restoration design (screwed vs. cemented) in reliability and failure modes of anterior crowns |
| |
Authors: | Freitas Amilcar C Bonfante Estevam A Rocha Eduardo P Silva Nelson R F A Marotta Leonard Coelho Paulo G |
| |
Affiliation: | Department of Biomaterials and Biomimetics, New York University College of Dentistry, New York, NY, USA. |
| |
Abstract: | The mechanical performance of cemented or screw-retained implant-supported crowns with an internal or external configuration is yet to be understood. This in vitro study evaluated the effect of screw-retained and cement-retained prostheses on internal and external implant-abutment connections. Thereby, the reliability and failure modes of crowns were investigated. Eighty-four implants (Emfils; Colosso Evolution system) were divided into four groups (n=21 each): screw-retained and internal connection (Si), screw-retained and external connection (Se), cement-retained and internal connection (Ci), and cement-retained and external connection (Ce). Ti-6Al-4V abutments were torqued (30 Ncm) to the implants, and maxillary central incisor metal crowns were torqued (30 Ncm) or cemented (Rely X Unicem; 3M-ESPE) and subjected to accelerated life-testing in water. Use-level probability Weibull curves and reliability for 50,000 cycles at 150 N were calculated. The β values for Si (1.72), Se (1.50), Ci (1.34), and Ce (1.77) groups indicated that fatigue/damage accumulation accelerated their failure. The Ci group presented the highest reliability, the Se group presented the lowest reliability, and Si and Ce groups presented intermediate reliability. Screw-retained restorations presented mainly abutment fracture. Cement-retained restorations resulted in failures of the screw in the Ce group, but implant/screw fracture in the Ci group. |
| |
Keywords: | cement‐retained fractography implant‐supported prostheses reliability screw‐retained |
本文献已被 PubMed 等数据库收录! |
|