首页 | 本学科首页   官方微博 | 高级检索  
     


Determination of impurities in heparin by capillary electrophoresis using high molarity phosphate buffers
Authors:Todd Wielgos  Karalyn Havel  Nadia Ivanova  Robert Weinberger
Affiliation:1. Baxter Healthcare Corporation, 25212 W. Illinois Route 120, Round Lake, IL 60073, United States;2. CE Technologies, Inc., PO Box 140, Chappaqua, NY, United States
Abstract:Oversulfated chondroitin sulfate (OSCS), an impurity found in some porcine intestinal heparin samples was separated from intact heparin by capillary electrophoresis (CE) using a 600 mM phosphate buffer, pH 3.5 as the background electrolyte in a 56 cm × 25 μm i.d. capillary. This method was confirmed in two separate labs, was shown to be linear, reproducible, robust, easy to use and provided the highest resolution and superior limits of detection compared to other available CE methods. Glycosoaminoglycans such as dermatan sulfate and heparan sulfate were separated and quantified as well during a single run. The heparin peak area response correlated well to values obtained using the official assay for biological activity. A high speed, high resolution version of the method was developed using 600 mM lithium phosphate, pH 2.8 in a 21.5 cm × 25 μm i.d. capillary which provided limits of detection for OSCS that were below 0.1%.
Keywords:Heparin   Oversulfated chondroitin sulfate   Heparin impurities   Capillary electrophoresis   Dermatan sulfate
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏