首页 | 本学科首页   官方微博 | 高级检索  
     


Efficacy of the laryngeal tube during intermittent positive-pressure ventilation
Authors:Asai T  Murao K  Shingu K
Affiliation:Department of Anesthesiology, Kansai Medical University, 10-15 Fumizono-cho, Moriguchi City, Osaka, 570-8507, Japan.
Abstract:We studied the efficacy of the laryngeal tube (VBM, Germany) during intermittent positive-pressure ventilation in 50 patients. After induction of anaesthesia and neuromuscular blockade, a size 4 laryngeal tube was inserted for patients of height 155 cm or greater. After insertion of the laryngeal tube, its pharyngeal and oesophageal balloons were inflated to an intracuff pressure of 60 cmH(2)O. An Ambu self-inflating bag was attached to the laryngeal tube and the lungs were ventilated manually at 15 breath.min(-1). It was possible to ventilate the lungs at the first attempt in 47 patients (94%). The airway pressure at which air leaked around the laryngeal tube exceeded 18 cmH(2)O in 41 patients (82%), and was > 30 cmH(2)O in 25 (50%). Median [interquartile range (range)] leak pressure was 30 [20-30 (6 to > 30)] cmH(2)O. Median [interquartile range (range)] tidal volume was 587 [533-653 (133-800)] ml or 8.8 [8.2-10.6 (1.9-12.6)] ml.kg(-1). We conclude that the laryngeal tube has a potential role in airway management during intermittent positive-pressure ventilation for anaesthesia or cardiopulmonary resuscitation.
Keywords:Ventilation    intermittent positive pressure    Airway
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号