首页 | 本学科首页   官方微博 | 高级检索  
     


Reliability of limb alignment measurement for high tibial osteotomy with a navigation system
Authors:J Lützner  AF Groß  KP Günther  S Kirschner
Affiliation:1. Department of Orthopaedic Surgery (Head of the department: Prof. Dr. K.-P. Günther), University Hospital Carl Gustav Carus, Medical Falculty of the Technical University of Dresden, Germany
Abstract:

Objective

High tibial osteotomy (HTO) is one treatment option for young and active patients with unicompartmental osteoarthritis. The success of this procedure substantially depends on the degree of correction of the mechanical axis. Computer-assisted navigation systems are believed to improve the precision of axis correction through intraoperative real-time monitoring. This study investigates the accuracy of limb alignment measurements with a navigation system on a cadaver specimen.

Materials and methods

The measurements were performed on a well-preserved cadaver specimen with a mechanical leg axis of 4° varus. Data was collected during the HTO workflow. Repeated serial measurements were undertaken by four different surgeons. After these measurements, different landmarks were deliberately set at the wrong place to examine the influence of mistakes during registration.

Results

There was a high intra-and interobserver reliability with a mean mechanical leg axis of 3.9° ± 0.7° and a mean error of 0.6°. The grossly incorrect placement of landmarks for knee and ankle center resulted in an incorrect mechanical leg axis of 1° valgus up to 10° varus.

Conclusion

The computer-assisted navigation system provided precise information about the mechanical leg axis, irrespective of the observer''s experience.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏