定位像扫描参数对胸部CT影像质量和辐射剂量影响的模体研究
引用本文:刘丹丹,崔莹,赵波,张永县,牛延涛. 定位像扫描参数对胸部CT影像质量和辐射剂量影响的模体研究[J]. 中华放射医学与防护杂志, 2021, 41(3): 217-221
作者姓名:刘丹丹  崔莹  赵波  张永县  牛延涛
作者单位:首都医科大学附属北京同仁医院放射科 100730
基金项目:国防科技创新项目(19-163-12-ZD-18-003-02);全军后勤科研项目(BKJ18J003)
摘    要:目的:探讨胸部CT定位像扫描参数(X射线管投照角度和管电压)的选择对图像质量和辐射剂量的影响规律,为临床实践中成像参数的选择提供指导。方法:选择不同扫描参数对成人胸部拟人模体进行定位像采集,X射线管投照角度(0°、90°和180°)、管电压(70、80、100、120和140 kV)和管电流(25 mA)共15种组合。...

关 键 词:X射线计算机体层成像  定位像  器官剂量  图像质量
收稿时间:2020-07-28

The influence of scout scanning parameters on image quality and radiation dose of chest CT:a phantom study
Liu Dandan,Cui Ying,Zhao Bo,Zhang Yongxian,Niu Yantao. The influence of scout scanning parameters on image quality and radiation dose of chest CT:a phantom study[J]. Chinese Journal of Radiological Medicine and Protection, 2021, 41(3): 217-221
Authors:Liu Dandan  Cui Ying  Zhao Bo  Zhang Yongxian  Niu Yantao
Affiliation:Department of Radiology, Beijing Tongren Hospital, Capital Medical University, Beijing 100730, China
Abstract:Objective To explore the effects of scout scanning parameters (projection angle and tube potential) on image quality and radiation dose in chest CT and to provide guidance for the selection of scanning parameters in clinical practice. Methods Different parameter combinations were selected to acquire the scout view of the anthropomorphic adult chest phantom. Fifteen scout views were obtained under parameter combinations of the projection angle (0/90/180 degrees), tube potential (70/80/100/120/140 kV) and tube current (25 mA). And then, chest spiral scans were performed according to the scout views. The other parameters were Assist kV, smart mA (maximum range), 80 mm detector width, 0.992:1 pitch, 0.5 s rotation time, 330 mm scanning length, 5 mm thickness, NI (noise index) 10. The radiation dose modulation index (ASiR-V) was 30%, and the iterative level was 50%. Volume CT dose index (CTDIvol) and tube current at 4 body levels (pulmonary apex, trachea bifurcation, mammary level and diaphragmatic dome) were recorded. The thermoluminescent dosimeters (TLDs) were used to measure the organ dose of the breast glandular. Regions of interest (ROIs)were selected at the trachea bifurcation and diaphragmatic dome to calculate the contrast noise ratio (CNR). Results The tube potential of the spiral scanning was automatically selected as 80 kV when the projection angle was 0 degree. The changing of the tube potential for the scout had little influence on the tube current at the four body levels, and the variation range was 0-2% (5/230). The tube potential in the spiral scanning was automatically selected as 100 kV when the projection angles were 90 and 180 degrees. The changing of the tube potential for the scout had a great influence on the tube current at the slice of trachea bifurcation and the variation range was from 14%(29/210)to 44%(93/210). According to the angle projection, there were statistical differences in CTDIvol(P<0.017), organ dose of the breast glandular, and CNR at the slices of trachea bifurcation and at diaphragmatic dome (F=13.027, 24.727, 10.630, P<0.05). According to the tube potential in the scout scanning, there were no significant differences in CTDIvol, organ dose of the breast glandular and CNR at both levels (P>0.05). Conclusions In chest CT, the influence of the projection angle in scout scanning on the image quality and radiation dose was more significant than that of the tube potential.
Keywords:Computed tomography  Scout view  Organ dose  Image quality