基于PET/CT后处理工作站行SPECT/CT骨定量分析的可行性
引用本文:余浩军,张一秋,李蓓蕾,石洪成.基于PET/CT后处理工作站行SPECT/CT骨定量分析的可行性[J].中国临床医学,2021,28(1):75-79.
作者姓名:余浩军  张一秋  李蓓蕾  石洪成
作者单位:复旦大学附属中山医院核医学科,复旦大学附属中山医院核医学科,复旦大学附属中山医院核医学科,复旦大学附属中山医院核医学科
基金项目:上海市卫生和计划生育委员会科研课题面上项目(201740201),上海市医苑新星青年医学人才培养(医学影像项目)资助计划(沪卫人事〔2020〕087号),上海市临床重点专科项目(shslczdzk03401)和西门子医疗科技有限公司资助项目(H2019-101)。
摘    要:目的:探讨单光子发射计算机断层成像术(SPECT/CT)和正电子发射断层成像术(PET/CT)后处理工作站对SPECT/CT骨定量分析数据进行定量分析结果的一致性,对比不同工作站测量病灶的最大标准摄取值(the maximum standardized uptake value,SUVmax)、平均标准摄取值(the ...

关 键 词:定量SPECT/CT  标准摄取值  "uWS-MI"医学影像处理软件
收稿时间:2020/5/31 0:00:00
修稿时间:2020/6/30 0:00:00

Feasibility study of SPECT/CT bone quantitative analysis using PET/CT post-processing workstation
YU Hao-jun,ZHANG Yi-qiu,LI Bei-lei,SHI Hong-cheng.Feasibility study of SPECT/CT bone quantitative analysis using PET/CT post-processing workstation[J].Chinese Journal Of Clinical Medicine,2021,28(1):75-79.
Authors:YU Hao-jun  ZHANG Yi-qiu  LI Bei-lei  SHI Hong-cheng
Institution:Department of Nuclear Medicine, Zhongshan Hospital, Fudan University,Department of Nuclear Medicine, Zhongshan Hospital, Fudan University,Department of Nuclear Medicine, Zhongshan Hospital, Fudan University,Nuclear Medicine Institute of Fudan University,Shanghai,;China;Shanghai Institute of Medical Imaging,Shanghai,;China
Abstract:Objective: To investigate the consistency of SPECT/CT and PET/CT post-processing workstations for quantitative analysis of SPECT/CT bone quantitative data, and compare the similarities and differences of the maximum standardized uptake value (SUVmax), the average standardized uptake value (SUVave), volume and average CT value of the lesions measured by different workstations.Methods: Forty patients who underwent whole-body bone scintigraphy and quantitative SPECT/CT imaging, and had no more than two hot spots within SPECT/CT imaging range were retrospectively analyzed. A total of 50 lesions were obtained. Three-dimensional (3D) isocontour volume of interest (VOI) method of "SyngoMI VB10B" SPECT/CT workstation was used to obtain the SUVmax, SUVave, volume, and average CT value of the lesions. Then the data were imported into the "uWS-MI" medical image processing software of PET/CT post-processing workstation. The above parameters were measured by 3D ellipsoid VOI method, adaptive threshold segmentation method, and fixed threshold segmentation method, respectively. The SUVmax, SUVave, volume, and average CT value measured by the three processing methods of "uWS-MI" workstation and 3D isocontour VOI method of "SyngoMI VB10B" SPECT/CT workstation were compared respectively to see if the difference was statistically significant.Results: The SUVmax measured by the 3D ellipsoid VOI method of "uWS-MI" workstation and the 3D isocontour VOI method of "SyngoMI VB10B" SPECT/CT workstation were completely consistent, but SUVave (t=9.841, P<0.001), lesion volume (t=-4.314, P<0.001) and average CT value (t=12.554, P<0.001) were significantly different. The adaptive threshold segmentation method and the fixed threshold method of "uWS-MI" workstation can not accurately segment two adjacent lesions of the lumbar spine and sacrum in one patient. Comparing with the 3D isocontour VOI method of "SyngoMI VB10B" SPECT/CT workstation, the SUVmax of 48 lesions in 39 other patients was completely the same. Comparing the adaptive threshold segmentation method of "uWS-MI" workstation with the 3D isocontour VOI method of "SyngoMI VB10B" SPECT/CT workstation, the SUVave (t=-0.723, P=0.473), lesion volume (t=-1.407, P=0.166), and average CT value (t=-0.735, P=0.466) of the 48 lesions were not statistically different. Comparing the fixed threshold measurement method of "uWS-MI" workstation with the 3D isocontour VOI method of "SyngoMI VB10B" SPECT/CT workstation, the SUVave (t=2.947, P=0.473) and lesion volume (t=-2.003, P=0.051) of 48 lesions were not statistically different, while the difference in average CT value (t=-4.71, P<0.001) was statistically significant.Conclusions: "uWS-MI" medical image processing software of PET/CT post-processing workstation has good compatibility with the quantitative data of SPECT/CT bone imaging. The three methods of PET/CT post-processing workstation can obtain consistent results for the measurement of SUVmax. However, the adaptive threshold segmentation method and fixed threshold measurement method of "uWS-MI" workstation have certain limitations for extracting the corresponding parameters for the segmentation of adjacent lesions. The values of SUVave, volume, and average CT value will be affected by different measurement methods. There is no significant difference between the adaptive threshold segmentation method of "uWS-MI" workstation and the 3D isocontour VOI method of "SyngoMI VB10B" SPECT/CT workstation.
Keywords:quantitative SPECT/CT  standardized uptake value  uWS-MI medical image processing software