塑料闪烁体探测器在光子束和电子束中的性能研究
引用本文:王美娇,姚凯宁,李莎,岳海振,刘卓伦,蒋璠,吴昊,王若曦. 塑料闪烁体探测器在光子束和电子束中的性能研究[J]. 中华放射医学与防护杂志, 2021, 41(3): 205-211
作者姓名:王美娇  姚凯宁  李莎  岳海振  刘卓伦  蒋璠  吴昊  王若曦
作者单位:北京大学肿瘤医院暨北京市肿瘤防治研究所放疗科 恶性肿瘤发病机制及转化研究教育部重点实验室 100142
基金项目:国家重大研发计划(2019YFF0104405;北京市自然科学基金(1202009)
摘    要:目的:探讨W2塑料闪烁体探测器在兆伏级光子束和电子束辐射中的性能。方法:采用直线加速器提供的光子束和电子束能量对W2闪烁体进行数据采集。研究内容包括静电计读数稳定性、W2剂量和剂量率线性以及角度响应,同时研究W2校准系数给剂量测量带来的不确定度。结果:静电计读数稳定性平均值的标准偏差在0.03~0.47之间;W2剂量的...

关 键 词:W2塑料闪烁体  切伦科夫校准因子  角度响应  放射治疗
收稿时间:2020-03-02

The performance of plastic scintillator detector in photon and electron beam
Wang Meijiao,Yao Kaining,Li Sh,Yue Haizhen,Liu Zhuolun,Jiang Fan,Wu Hao,Wang Ruoxi. The performance of plastic scintillator detector in photon and electron beam[J]. Chinese Journal of Radiological Medicine and Protection, 2021, 41(3): 205-211
Authors:Wang Meijiao  Yao Kaining  Li Sh  Yue Haizhen  Liu Zhuolun  Jiang Fan  Wu Hao  Wang Ruoxi
Affiliation:Key Laboratory of Carcinogenesis and Translational Research (Ministry of Education/Beijing), Department of Radiotherapy, Peking University Cancer Hospital & Institute, Beijing 100142, China
Abstract:Objective To investigate the performance of W2 plastic scintillator in megavolt photon and electron beams. Methods The photon and electron beam energy provided by linear accelerator was used to collect data of the W2 scintillator. The parameters include the electrometer reading stability, W2 dose and dose rate linearity, and angular response. And the dose uncertainty of the W2 correction factors was also investigated. Results The standard deviation of the electrometer reading stability was between 0.03 and 0.47. The linear regression factors of W2 dose were all 1.0; the maximum deviation of the dose rates was 0.61%. The Cerenkov light radiation correction factor(CLR) for 6 and 10 MV were 0.741 and 0.746, respectively, and the CLR for 6, 9, 12 and 15 MeV were 0.750, 0.753, 0.757 and 0.757, respectively. The maximum deviation of dose uncertainty for 15 MeV was 3.15%. Conclusions The signal obtained by the blue and green channel was no angular dependence, the same as the high-energy electron beam, which verified that the Cerenkov radiation correction factor has good linearity. W2 plastic scintillator can be applied to non-coplanar radiotherapy dosimetry.
Keywords:W2 plastic scintillator  Cerenkov light radiation correction factor  Angular response  Radiation therapy