DDDR起搏器程控功能的临床应用价值
引用本文:周金台,万征,刘维宇,王志毅,仉玉书,李琼.DDDR起搏器程控功能的临床应用价值[J].中国心脏起搏与心电生理杂志,1995(2).
作者姓名:周金台  万征  刘维宇  王志毅  仉玉书  李琼
作者单位:天津医科大学总医院
摘    要:对埋植DDDR起搏器的21例患者随访26(6~54)个月,未见电极移位。20条激素电极慢性能量阈值为0.44±0.51μJ,有意义地低于15条螺旋电极的1.89±1.98μJ。通常程控到慢性能量阈值的3倍以获得适当的安全范围,其中65%的激素电极程控到1.6V,0.3ms仍具3.94倍能量阈值的安全范围。提示具有较大范围可程控电压和脉宽输出的DDDR起搏器配用低阈值激素电极可延长起搏器寿命,这也将是今后起搏系统设计的方向。

关 键 词:DDDR起搏器,程控,寿命,起搏阈值,电极

Programming and Longevity of DDDR Pacemaker
Zhou Jintai,Wan Zheng,Liu Weiyu,et al.Programming and Longevity of DDDR Pacemaker[J].Chinese Journal of Cardiac Pacing and Electrophysiology,1995(2).
Authors:Zhou Jintai  Wan Zheng  Liu Weiyu  
Abstract:The longevity of physiological and safe pacing was evaluated in 21 patients implanted with a DDDR pacemaker of advanced technology. No dislodgement of lead was observed during a follow-up of 6~54(average 26 ) months.The chronic energy threshold in 20 steroid leads was 0.44 ± 0.51μJ and was significantly lower than in 15 screw-in leads(1.89±1.98μJ). For an adequate safety margin, the pacing outputs were programmed to 3 times the chronic energy threshold and even a 1.6 V,0.3ms pacing output which offered a 3.94 times safety margin.Our experience suggests that a DDDR pacemaker with a greater range of output programmability(voltage and pulse width) plus two low threshold steroid electrodes can prolong the unit life span and will be an objective of future technology design of pacing system.
Keywords:DDDR pacemaker  Programming  Longevity  Pacing threshold  Electrode  