Intraoperative awareness due to malfunction of a Siemens 900B ventilator
Authors:Slinger  Peter D.  Andrew  W.  Scott  C.  Kliffer  A. Paul
Affiliation:Department of Anaesthesia, Montreal General Hospital, Quebec, Canada.
Abstract:A case of intraoperative awareness during a thoracotomy is described. The patient's recall coincided with an intraoperative period during which a Siemens 900B ventilator and a Siemens 952 isoflurane vaporiser were used. Subsequent assessment of this equipment with an anaesthetic agent analyzer revealed that, at the ventilator settings which had been used, the delivered anaesthetic vapour concentration varied greatly from the vaporizer settings. This problem eventually was traced to a malfunctioning inlet control valve on the ventilator. This complication may have been prevented if the end-tidal anaesthetic concentration had been monitored intraoperatively.
Keywords:  font-variant:small-caps"  >anaesthesia: depth, awareness     font-variant:small-caps"  >complications: awareness     font-variant:small-caps"  >equipment: ventilators, malfunction