Determination of airway humidification in high-frequency oscillatory ventilation using an artificial neonatal lung model |
| |
Authors: | H. Schiffmann S. Singer D. Singer E. v. Richthofen J. Rathgeber K. Züchner |
| |
Affiliation: | Department of Pediatrics, University of G?ttingen, Robert-Koch-Strasse 40, D-37 073 G?ttingen, Germany Tel.: + 49 (551) 396222 Fax: + 49 (551) 396231 email: schiffi@med.uni-goettingen.de, DE Department of Anesthesiology, University of G?ttingen, G?ttingen, Germany, DE
|
| |
Abstract: | Objective: Thus far only few data are available on airway humidification during high-frequency oscillatory ventilation (HFOV). Therefore, we studied the performance and efficiency of a heated humidifier (HH) and a heat and moisture exchanger (HME) in HFOV using an artificial lung model. Methods: Experiments were performed with a pediatric high-frequency oscillatory ventilator. The artificial lung contained a sponge saturated with water to simulate evaporation and was placed in an incubator heated to 37 °C to prevent condensation. The airway humidity was measured using a capacitive humidity sensor. The water loss of the lung model was determined gravimetrically. Results: The water loss of the lung model varied between 2.14 and 3.1 g/h during active humidification; it was 2.85 g/h with passive humidification and 7.56 g/h without humidification. The humidity at the tube connector varied between 34.2 and 42.5 mg/l, depending on the temperature of the HH and the ventilator setting during active humidification, and between 37 and 39.9 mg/l with passive humidification. Conclusion: In general, HH and HME are suitable devices for airway humidification in HFOV. The performance of the ventilator was not significantly influenced by the mode of humidification. However, the adequacy of humidification and safety of the HME remains to be demonstrated in clinical practice. Received: 22 September 1998 Final version received: 22 March 1999 Accepted: 23 April 1999 |
| |
Keywords: | Measurement techniques Respiration Airway humidification methods |
本文献已被 SpringerLink 等数据库收录! |
|