Performance characteristics of a new less sensitive HIV-1 enzyme immunoassay for use in estimating HIV seroincidence |
| |
Authors: | Kothe Denise Byers Robert H Caudill Samuel P Satten Glen A Janssen Robert S Hannon W Harry Mei Joanne V |
| |
Affiliation: | Division of Laboratory Sciences, National Center for Environmental Health, Atlanta, Georgia 30341, USA. |
| |
Abstract: | Less sensitive (LS) HIV-1 enzyme immunoassays (EIAs) have significantly improved the quantity and quality of HIV surveillance data. The first LS-HIV-1 EIA, the Abbott 3A11-LS, provided reliable incidence data, but the assay required specialized equipment, and the lack of available reagents made testing difficult. This study evaluated the use of an alternate assay, a modified version of the Vironostika HIV-1 EIA (Vironostika-LS), to be used for LS testing. The Vironostika-LS has similar performance characteristics to the Abbott 3A11-LS with additional advantages. This 96-well formatted assay is commonly found in public health laboratories for routine HIV-1 testing and can be used with both serum and dried blood spot specimens. The estimated mean time from seroconversion (defined using a standardized optical density cutoff of 1.0) with the Vironostika-LS was 170 days (95% CI, 145-200 days). When the Vironostika-LS was applied to a matched serum set previously tested with the Abbott 3A11-LS, the Vironostika-LS accurately identified 97% of specimens with recent or long-standing HIV infection. The paper also reports Vironostika-LS quality control guidelines and the results from 3 rounds of proficiency testing. |
| |