A New Tracheostomy Tube
Authors:N. Lomholt  M.D.  S. Borgeskov   B. Kirkby
Affiliation:Department of Pharmacology, University of Copenhagen and Department of Head and Neck Surgery, University Hospital, Copenhagen, Denmark
Abstract:Recent publications show that severe damage to the trachea is still a problem with high-volume, low-pressure cuffs. The NL tracheostomy tube was used in 85 patients for 3 days to 2 months (mean 16 days). This tube has a high-volume, low-pressure cuff with automatic regulation of the cuff pressure at 3kPa. The tube has a flexible tip. Fiberoptic examination at extubation showed minimal damage to the tracheal mucosa: 33 patients had normal mucosa and the rest had hyperaemia and/or fibrin formation. Four patients had ulcerations from suction catheters and four patients had small, superficial ulcerations produced by the tip of the lube. Of these last four patients, three had skin flaps that exerted pressure on the tube. Severe tracheal damage was prevented due to the combination of automatic regulation of cuff pressure and a flexible tip of the tube.