容积型指脉波图监测及评价体外反搏效果的对照研究
引用本文:伍贵富,杨世方.容积型指脉波图监测及评价体外反搏效果的对照研究[J].生物医学工程学杂志,1999,16(4):493-496.
作者姓名:伍贵富  杨世方
作者单位:中山医科大学附属第一医院心脏内科!广州510080,中山医科大学附属第一医院心脏内科!广州510080,中山医科大学附属第一医院心脏内科!广州510080,中山医科大学附属第一医院心脏内科!广州510080,中山医科大学附属第一医院心脏内科!广州510080,中山医科大学附属第一医院心脏内科!广
摘    要:对20 名男性健康志愿者采用经桡动脉穿刺置管主动脉根部测压的方法,分别研究不同反搏工作参数对主动脉压力和指脉波的影响。结果反搏时主动脉舒张压随储气罐压力的增加升高(最高可达168.56 m m Hg);主动脉舒张波的dp/dt值均逐渐上升,但收缩波的dp/dt值则呈下降趋势;指脉舒张波起点与主动脉舒张波起点比较平均滞后(108.20±25.70) m s,而指脉收缩波起点则平均滞后(96.69±38.70) m s;指脉波的D/S比值与主动脉压力波的D/S比值随反搏储气罐压力的增加而同步增加,并由直线回归分析得出动脉D/S比值(Y)与指脉D/S比值(X)的回归方程式Y= 0.376+ 0.457X。本研究指示体外反搏对动脉压力波的D/S比值、舒张压及其dp/dt值均有显著影响,指脉波仍然是目前调节反搏工作参数及监测反搏效果的较为安全、可靠的方法

关 键 词:体外反搏  脉波图  血压  监测

A Comparative Study of Finger Plethysmography and Aortic Pressure for Monitoring the Effect of External Counterpulsation
Wu Guifu Zheng Zhensheng Du Zhimin Zhang Miaoqing,Hu Chengheng,Liao Ying,Zhao Shunqing,Yang Shifang,Wu Fujun.A Comparative Study of Finger Plethysmography and Aortic Pressure for Monitoring the Effect of External Counterpulsation[J].Journal of Biomedical Engineering,1999,16(4):493-496.
Authors:Wu Guifu Zheng Zhensheng Du Zhimin Zhang Miaoqing  Hu Chengheng  Liao Ying  Zhao Shunqing  Yang Shifang  Wu Fujun
Institution:Department of Cardiology, First Affiliated Hospital, Sun Yat-sen University of Medical Sciences, Guangzhou 510080.
Abstract:Invasive pressure detecting and monitoring of aortic root was conducted in 20 volunteers so as to study the effect of Enhanced External Counterpulsation (EECP) on aortic pressure and finger plethysmography. The results demonstrated that the changes of aortic diastolic pressure linearly related to the counterpusation's bladder pressure, and the aortic augmentation increased to the top of 168.56 mmHg with a D/S ratio of 2.75(amplitude ratio of diastolic waveform and systolic waveform, D/S ratio) following the bladder pressure of 0.45 Kg/cm2. With the increase of bladder pressure, the dp/dt of aortic augmentation elevated gradually, but a down-regulating trend was seen in systolic waveform. During EECP, the starting time of plethysmographic augmentation had a mean delay of 108.20 +/- 25.70 ms comparing with that of aortic augmentation. The duration between the starting time of aortic systolic waveform and that of finger plethysmogram was arround 96.69 +/- 38.70 ms. When the bladder pressure changed, no significant difference was observed among them. The D/S ratio of finger plethysmographic waveform changed in concord with that of aortic pressure with the promotion of bladder pressure. Therefore, a relationship between aortic D/S ratio (Y) and plethysmographic D/S ratio (x) was constructed by the linear regression analysis: Y = 0.376 + 0.457X. It can be concluded that EECP has significant effects on aortic D/S ratio, aortic peak augmentation and its dp/dt value. Finger plethysmography, to date, still plays important roles in monitoring and controlling EECP work safely and effectively.
Keywords:External counterpulsation  Plethysmography  Blood pressure  