The pericardial valve in the aortic position ten years later |
| |
Authors: | L Gonzalez-Lavin J Gonzalez-Lavin S Chi B Lewis S Amini D Graf |
| |
Institution: | Department of Surgery, Robert Wood Johnson University Hospital, New Brunswick, N.J. |
| |
Abstract: | To assess the behavior of the pericardial valve at 10 years after implantation, the cases of 240 patients who had undergone aortic valve replacement with the standard Ionescu-Shiley (Shiley, Inc., Irvine, Calif.) bovine pericardial valve between February 1977 and December 1983 were reassessed. Follow-up of the 224 hospital survivors was 99.6% complete. Fifty-seven valve-related events occurred. Fourteen were thrombotic events (1.2%/patient-year), 28 were intrinsic tissue failures (2.4%/patient-year), 13 were cases of prosthetic valve endocarditis (1.1%/patient-year), and 2 were paravalvular leaks (0.17%/patient-year). The linearized rate for death, reoperation, or both resulting from valve-related events was 3.6%/patient-year. Time-related hazard function for the instantaneous risk of death and/or reoperation resulting from valve-related events demonstrated an exponential increase after 80 months. These data, in conjunction with our previous reports on the histologic changes in pericardial collagen and the incidence of calcification (26/28), should be considered regarding new and future generations of pericardial bioprostheses. Although this device provides good hemodynamics and carries a low incidence of thromboembolism, it has a limited durability. New generations of pericardial valves may have improved structural features, but the behavior of glutaraldehyde-fixed, formaldehyde-stored bovine pericardium as currently selected and prepared is unlikely to change. |
| |